Chapter 2. Installation
16
PanaFlow™ LC User’s Manual
3. Determine the upstream and downstream ends of the pipe and place the appropriate transducer into the
corresponding block on one of the sub-assemblies. Make sure the transducer cable connector faces away from
the center of the installation.
Figure 20: Transducer Mounting on UCF (Top Left), MCF (Top Right), GCF (Bottom Left), and SCF (Bottom Right)
4. Use the pressure bolt to secure the transducer in place. The pressure bolt should fit into the dimple on the
transducer. Hand tighten the bolt just enough to hold the transducer in place, but do not overtighten it or the
fixture will lift off the pipe.
5. Repeat Steps 1-4 to mount the other transducer in the fixed short block. See
clamping fixture installations.
6. Connect the other end of the transducer cables to the junction box in Section 4: Transducer Wiring of this manual.
Note:
If the transducers are mounted into the clamping fixtures properly, the two transducer cable connectors will
face away from each other as shown above.