PanaFlow™ LC User’s Manual
11
Chapter 2. Installation
Figure 10: SCF Odd Number of Traverse Installation, Step 4
4. From one of the marks on the top of the pipe, measure around the pipe a distance equal to 1/4 of the pipe
circumference, or a distance that will satisfy the orientation found in Step 1. Use the center punch to make a mark
at this point.
Figure 11: SCF Odd Number of Traverse Installation, Step 5
5. From the other mark on the top of the pipe, measure around the pipe in the opposite direction a distance equal
to 1/4 of the pipe circumference, or the same distance used in Step 5. Use the center punch to make a mark at
this point.
Figure 12: SCF Odd Number of Traverse Installation, Step 6
6. Center one of the blocks over one of the center punch marks on the side of the pipe. Align the block so that the
pressure bolt is directly over the punch mark. Secure the block by wrapping the two steel straps around both the
block and the pipe and tightening the straps.
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...panametrics com BH029C11 EN C PanaFlow LC User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 64: ...Appendix D CE Mark Compliance 56 PanaFlow LC User s Manual no content intended for this page...
Page 66: ...Warranty 58 PanaFlow LZ User s Manual no content intended for this page...
Page 67: ......