PanaFlow™ LC User’s Manual
9
Chapter 2. Installation
2.4.3.1
Install of SCF Fixture (Even Traverses)
Note:
The instructions in this section can also be used for a multiple-traverse method. However, you must use an
EVEN
number of traverses. The distance the signal travels from one side of the pipe wall to the opposite side
of the pipe wall is considered one traverse. For installations with more than two traverses, contact
Panametrics for assistance.
The installation procedure for transducers using the even number of-traverse method is as follows:
1.
Prepare the pipe where you intend to place the clamping fixture by making sure it is clean and free of loose
material. Sanding, though usually not required, may be necessary to remove any high spots. However, be careful
to preserve the original curvature of the pipe.
2. Secure one of the blocks to the alignment bar by tightening one of the nuts on the block so the block will not be
able to slide along the alignment bar's slot.
3. Obtain the transducer spacing dimension (
S
) after programming the PanaFlow LC transmitter. Using the
alignment bar as a guide, move the second block so that the distance between the blocks equals the S
dimension. Use the pressure bolts or the edges of the blocks as reference points, as shown below
Figure 7: SCF Setup, Spacing
4. Position the clamping fixture on top of the pipe and wrap the clamping fixture blocks with the hose clamps.
Tighten the clamps until the blocks are snug but ensure that they can still be rotated around the pipe.
5. Rotate the blocks so that they are away from the vertical plane and in an orientation that satisfies the criteria
listed in Section 2.6:
Even and Odd Number of Traverse Installation
.
6. Fully tighten the hose clamps so no movement on the fixture will occur.
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...panametrics com BH029C11 EN C PanaFlow LC User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 64: ...Appendix D CE Mark Compliance 56 PanaFlow LC User s Manual no content intended for this page...
Page 66: ...Warranty 58 PanaFlow LZ User s Manual no content intended for this page...
Page 67: ......