Invivo
2
Plus
UM-14 Version R3.0
Page 10
Invivo
2
Plus User Manual
Gas and electrical supply requirements
2.
Gas supply requirements
2.1
The Invivo
2
Plus workstation requires;
•
Oxygen free nitrogen.
•
In addition to this, the Invivo
2
Plus workstation should be connected to ICO
2
N
2
IC (supplied
with the Invivo
2
Plus workstation). All other gas connections should be made to the
ICO
2
N
2
IC.
See UM-011 ICO
2
N
2
IC user manual for more details.
The maximum permissible concentration of hydrogen in the anaerobic mixed gas is 5.5%. The use
of any other gas(ses) will invalidate the warranty and may cause serious injury. The Gas supplies
must be securely fastened after installation.
Gas regulator requirements
2.2
Gas regulators must be used with all gas supplies to the workstation. A 2 stage regulator must be
used for each bottled gas supply. Regulators should be available from your local gas supplier.
Ruskinn Technology Limited does not supply gas regulators.
The gas regulator should provide a minimum supply pressure of 3 (42psi) Bar Gauge. The maximum
supply pressure permissible is 4 bar gauge. A supply pressure greater than this will damage internal
components of the workstation and will invalidate the warranty.
The gasses are connected to the ICO
2
N
2
IC Advanced Gas mixing system. See UM-011 ICO
2
N
2
IC user
manual for more details.
Electrical supply requirements
2.3
The workstation must be connected to a mains power supply. A power cord is supplied to connect
the workstation to the mains supply. Only the power cord supplied should be used to connect the
workstation to the mains supply. It is advised that the workstation be located no greater than 1
metre from the plug socket. The plug socket should not be obscured by the Invivo
2
Plus workstation.
The workstation must be connected to a protective earth.
NOTE:
If an Uninterruptable Power Supply (UPS) system is used, both the Invivo
2
Plus workstation
and the ICO
2
N
2
IC Advanced Gas mixing system must be connected to the Uninterruptable Power
Supply (UPS) system.
Voltage and frequency requirements
2.4
To ensure safe operation of the workstation, it must be connected to a supply of the correct voltage
and frequency, as shown in the rating label (Figure 8). The mains supply voltage fluctuations must
not /- 10% of the nominal mains voltage.
Power consumption
2.5
Figure 3: Power consumption ratings
Supply voltage and frequency
Power consumption
240V AC, 50Hz
200W
120V AC, 60Hz
250W