Warnings
There are various warning screens and icons to draw the user attention when required.
Temperature Alarm
This screen will appear if the workstation temperature is not at the set-point. This will
appear during the initial turn-in or if the set-point has been changed by more than 0.2°C.
The alarm can be dismissed by pressing the return arrow at the bottom right of the screen.
If the workstation still has not reached set-point within 20 minutes then the alarm screen
will return. If the alarm screen is not dismissed manually it will automatically be cancelled
if the workstation reaches set-point.
Screen 20: Temperature Alarm Screen