Building exhaust requirement 

Your ChemoSHIELD

®

 cabinet has an 8” exhaust collar and requires the building exhaust connection to 

handle 1.8” WC at 310 CFM. 

 

Chamber walls and tunnel construction 

The chamber walls and tunnel of your ChemoSHIELD

®

 are constructed of 16 gauge, type 316 stainless 

steel.  The easy-to-clean 7/16-inch radiused (rounded) corners prevent buildup of contaminants and resist 
corrosion. The chambers are designed to be leak tight to 2” WC. 

 

All-metal plenums 

The plenums of your ChemoSHIELD

®

 unit have been constructed entirely of metal in order to provide 

strength, durability, air-tightness and resistance to deterioration. 

 

Stainless steel work surfaces 

The work surfaces are constructed of corrosion-resistant 16-gauge, type 316 stainless steel. They have a #4 

satin finish that diminishes light reflection.  

The work chamber has a three (3) piece flat work surface and the pass-through chamber has a two (2) 

piece flat work surface. The work surfaces are removable so that you can access the exhaust filters. 

 

Cabinet exterior panels 

External panel construction is of 16-gauge cold-rolled steel protected by a smooth white baked enamel 

finish. 

 

Tested HEPA filters 

The filters in the ChemoSHIELD

®

 are scan-tested HEPA filters.  They are 99.99% effective on particles of 

0.3 micron in size. Each filter is scan-tested after installation to assure that there are no leaks >.01% of 
upstream concentration. 

 

Viewscreens  

Your ChemoSHIELD

®

 has 3/8-inch shatterproof, clear polycarbonate in both the pass-through door and 

the work chamber viewscreen. 

 

Work chamber and pass-through lighting 

The lighting system produces at the work surface of the work chamber an average of 50 foot-candles of 

illumination and 45 foot-candles at the work surface of the pass-through chamber. Two fluorescent lamps 
are externally mounted to minimize heat buildup.  The cabinet uses one F39T5/HO fluorescent lamp for the 
work chamber and one FT36W-2G11-41 fluorescent lamp for the pass-through chamber. 

 

Chamber base pan construction and clean ability 

The base pan is constructed of 16-gauge, type 316 stainless steel with 7/16” radius in all corners to 

facilitate cleaning. The work surfaces can be repositioned for cleaning of the chamber pans. 

 

Cabinet air balance adjustments 

Only a technician with proper training and equipment should perform air balancing.  (See Section IV, "On-

Site Checks and Maintenance Procedures.") 

351D700 Rev G 

 

 

 

 

Summary of Contents for ChemoSHIELD CS500

Page 1: ...o bakerco com bakerco com OPERATOR S MANUAL Negative Pressure Isolator MODEL CS500 Revision Date Description H 5 19 15 UPDATED BRANDING This manual includes information for installation operation main...

Page 2: ...laboratory personnel In addition to the high quality you expect from all Baker equipment this model has been ergonomically designed to provide the lab user with many exciting design features The ergo...

Page 3: ...3 Easy filter access 4 Telescoping stand 4 Mechanical interlock between pass through door inter chamber door 4 Pressure monitoring of chambers 4 Airflow monitor AFM 4 Auxiliary switch Optional 4 Elec...

Page 4: ...HEPA FILTERS 12 REMINDERS 12 APPENDIX 13 REPLACEMENT PARTS LIST 14 AIRFLOW MONITOR OPERATION CALIBRATION 15 WARRANTY 16 WARNING The agents used and or work performed in this cabinet may be hazardous t...

Page 5: ...ection while the sealed work chamber and HEPA filtered airflow provides product protection Fig 1 Airflow Inside Cabinet Negative Pressure The pass through chamber allows operators to introduce or remo...

Page 6: ...ny features that sets the ChemoSHIELD unit apart from other cabinets is the interaction of its pressure areas The whole cabinet is under negative pressure when operating The pass through chamber opera...

Page 7: ...protected by a smooth white baked enamel finish Tested HEPA filters The filters in the ChemoSHIELD are scan tested HEPA filters They are 99 99 effective on particles of 0 3 micron in size Each filter...

Page 8: ...are comparing chamber pressure to the room pressure Airflow monitor AFM The AFM is an audible and visual alarm system consisting of solid state flow measuring and alarm circuitry in a small enclosure...

Page 9: ...at the site of installation The cabinet is to be powered by a 115 V AC 15A 60 Hertz Single Phase dedicated circuit This circuit should be protected by a 15A circuit breaker or fuse Typical Current for...

Page 10: ...ndoor use Altitudes up to 2000 meters Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Main supply voltage 105 to 125Volts AC Transient over...

Page 11: ...Explosive or flammable substances should never be used in this cabinet Location Within the Laboratory Position the ChemoSHIELD so the pass through chamber end is 15 away from any wall or obstruction...

Page 12: ...guidelines Reference the cabinet factory test report for information on the test criteria used NOTE This test should be performed before the cabinet is connected to the building exhaust system See Fig...

Page 13: ...ch will illuminate when the fluorescent lights are turned on 2 Auxiliary AUX Switch Optional This switch is available for wiring to a building exhaust system fan blower so that it can be turned on off...

Page 14: ...been open and closed to allow the chamber and tunnel to purge with HEPA filtered air Unloading and Wipe down 1 Close the inter chamber door before opening the pass through door when unloading material...

Page 15: ...m being killed Standard disinfectants include Hypochloride chlorine bleach Iodophor Detergent Ethanol Phenol and Alcohol IMPORTANT Disinfection and cleaning should always be followed by rinsing in ste...

Page 16: ...Company for more information It must be pointed out that the HEPA filter is not intended to filter gasses or vapors nor are they 100 efficient on particulates The room and cabinet particulate levels...

Page 17: ...Appendix 351D700 Rev G 13...

Page 18: ...13 x 6 39797 Filter HEPA 9 x 13 x 6 40817 Replacement Filters kit for complete cabinet 351G000 Latch Inter Chamber Door 40114 Latch Pass Through Door 40114 Latch Work Chamber Viewscreen 40881 Nitrile...

Page 19: ...n normal airflow is resumed the BLOWER relay will change states and will have a slower single audible alarm The red LED will be latched ON solid until reset with the Alarm Reset button No Probe Red an...

Page 20: ...nc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any incidental or consequenti...

Reviews: