Transferring material into the workstation chamber via the interlock
To transfer material into the workstation chamber via the interlock;
•
Place the Petri dishes in a large Petri dish holder.
•
Open the interlock outer door (see section 8.1.2 for more details).
•
Place the Petri dish holder inside the interlock.
•
Close the interlock outer door (see section 8.1.2 for more details).
•
Start the interlock purge cycle by pressing the interlock purge button on the touchscreen
•
Access the workstation chamber via the Ezee Sleeves (see section 8.2 for more details).
•
Open the interlock inner door (see section 8.1.3 for more details).
•
Remove the Petri dish holder from the interlock.
•
Close the interlock inner door (see section 8.1.3 for more details).
Removing material from the workstation chamber via the interlock
To remove material from the workstation chamber via the interlock;
•
Open the interlock inner door (see section 8.1.3 for more details).
•
Place the Petri dish holder in the interlock.
•
Close the interlock inner door (see section 8.1.3 for more details).
•
Exit the workstation chamber via the Ezee Sleeves.
•
Open the interlock outer door (see section 8.1.2 for more details).
•
Remove the Petri dish holder from the interlock.
•
Close the interlock outer door (see section 8.1.2 for more details).