Transport and storage
When not in use, the Bugbox Ax Workstation must only be stored within a temperature of between
0°C and 30°C
Storage outside of this range may damage the workstation.
Location and handling of the Bugbox Ax workstation
The Bugbox Ax should only be installed or relocated by a qualified engineer. To arrange
installation or relocation please contact your local distributor.
The mains appliance coupler and plug are the AC mains supply isolation device and must be
easily accessible.
For optimal performance, the Bugbox Ax should not be installed directly beneath Air Conditioning
outlets. This can cause excessive humidity to collect on the inside surface of the front screen.
Environmental Operating Conditions
The Bugbox Ax Workstation should only be operated under the following environmental
conditions:
•
Temperature
–
Between 15
⁰
C and 30
⁰
C
•
Humidity
–
Between ambient and 90% RH, Non-Condensing
The workstation must be located in a well-ventilated area.