Page 34 of 50
Warnings
There are various warning screens and icons to draw the user attention when required.
Temperature Alarm
This screen will appear if the workstation temperature is not at the set-point. This will
appear during the initial turn-in or if the set-point has been changed by more than 0.2°C.
The alarm can be dismissed by pressing the return arrow at the bottom right of the screen.
If the workstation still has not reached set-point within 20 minutes then the alarm screen
will return. If the alarm screen is not dismissed manually it will automatically be cancelled
if the workstation reaches set-point.
Screen 13: Temperature Alarm Screen