28
5.3 Configuring AutoCal
Use the instructions in this chapter to configure the AutoCal process as desired. Begin by navigating to the
cal>1:cal>AutoCal
menu. Then, proceed to the next section to begin.
5.3.1 Enabling AutoCal
The first step in configuring AutoCal is to program whether
this process is activated or deactivated. To perform this task,
proceed as follows:
1. At the
AutoCal
menu, press [ENTER].
2. Use the [
] and [
] keys to highlight the
enable/disable
option and press [ENTER].
3. Use the [
] and [
] keys to highlight either the
enable
option or the
disable
option and press [ENTER].
4. If you chose to disable AutoCal, no additional AutoCal
programming is required. If you chose to
enable
AutoCal,
continue the programming in the next section.
5.3.2 Selecting the AutoCal type
After enabling AutoCal, the next step is to select the type
of calibration to perform. The following three options are
available:
•
Offset calibration
- this is a one-point calibration that
uses an offset correction at one specified ppm/pcnt value.
•
Zero/span calibration
- this is a two-point calibration that
uses an offset correction at both the zero and span (full-
scale) ppm/pcnt values.
NOTE:
The instructions in this manual assume that the
XDP has been programmed to calibrate the zero point
first (see
“zero/span cal type”
on page 56
). If your unit is
configured to calibrate the span point first, the option
above will be shown as “span/zero calibration” and steps
5a and 5c on the next page will be reversed.
•
Slope calibration
- this is a one-point calibration that
uses an offset correction at just the span (full-scale)
ppm/pcnt value.
NOTE:
The
slope
calibration is really just a special case
of the zero/span calibration where there is no offset
correction at the zero point.
To select the AutoCal type, proceed as follows:
1. Enable
AutoCal
as described in the previous section.
2. Use the [
] and [
] keys to highlight the
AutoCal types
menu option and press [ENTER].
3. Use the [
] and [
] keys to highlight the desired
calibration method and press [ENTER].
Proceed to the appropriate step below for the type of
calibration you chose in step 3 above.
4. For the
offset cal
method, proceed as follows:
a. Using the four arrow keys, program the ppm/pcnt
value that you wish to use for the offset correction.
b. Press [ENTER] to complete the programming.
5. For the
zero/span cal
method, proceed as follows:
a. Using the four arrow keys, program the ppm/pcnt
value that you wish to use for the zero correction.
b. Press [ENTER].
c. Using the four arrow keys, program the ppm/pcnt
value that you wish to use for the span (full-scale)
correction.
d. Press [ENTER] to complete the programming.
6. For the
slope cal
method, proceed as follows:
a. Using the four arrow keys, program the ppm/pcnt
value that you wish to use for the span (full-scale)
correction.
b. Press [ENTER] to complete the programming.
5.3.3 Setting calibration gas flow times
To optimize the AutoCal process, you may program
the length of the gas flow times during a calibration.
To do so, proceed as follows:
1. Enable
AutoCal
as described on page 5-2.
2. Use the [
] and [
] keys to highlight the
AutoCal control
menu option and press [ENTER].
3. Use the [
] and [
] keys to highlight the
zero gas on time
option and press [ENTER].
4. Using the four arrow keys, program the time in minutes
that you wish to allow for the zero gas to settle.
Press [ENTER].
5. Use the [
] and [
] keys to highlight the
span gas on
time
option and press [ENTER].
6. Using the four arrow keys, program the time in minutes
that you wish to allow for the span gas to settle.
Press [ENTER].
7. Use the [
] and [
] keys to highlight the
gas settling time
option and press [ENTER].
8. Using the four arrow keys, program the time in minutes
that you wish the allow for the process gas to settle.
Press [ENTER].
9. Use the [
] and [
] keys to highlight the
gas measuring
time
option and press [ENTER].
10. Using the four arrow keys, program the time in minutes
that you wish to allow for the data collection. Then,
press [ENTER].
Summary of Contents for Panametrics XDP
Page 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 8: ...viii no content intended for this page...
Page 12: ...no content intended for this page proceed to next page 2...
Page 28: ...no content intended for this page proceed to next page 18...
Page 36: ...26 no content intended for this page proceed to next page...
Page 42: ...32 no content intended for this page proceed to next page...
Page 58: ...48 no content intended for this page proceed to next page...