oxy.IQ User’s Manual
11
Chapter 3. Initial Setup and Operation
3.3.2.3
Trimming the Analog Output High (20 mA) End
1.
Press the
key and then press the
Enter
key to enter the 20 mA
Trim
menu, and the analog output is driven
to about 20 mA.
2.
Use the
and
keys to adjust the analog output up or down, until it equals 20.00 ± 0.01 mA.
3.
Press the
Enter
key to save the trim adjustment and return to the
Trim
menu.
3.3.2.4
Completing the Trim Procedure
1.
Repeat both the low (4 mA) end and high (20 mA) end analog output trimming steps until no further trimming
adjustments are required.
2.
Press the
Cancel
key twice to return to the
Main Menu
.
3.3.3 Air Calibration
An air calibration is always recommended upon installation of a new oxygen sensor. However, because of the
non-linearity of the oxygen sensor, a span gas calibration (see the next section) should also be performed to ensure
more accurate readings in the ppm ranges.
To perform an air calibration, complete the following steps:
1.
Press the
Enter
key to enter the
Main Menu
.
2.
Press the
Enter
key to enter the
Calibration
menu.
3.
Press the
Enter
key to select the
Air
menu option.
Proceed to the appropriate section, depending on whether you are calibrating a new sensor or recalibrating an
existing sensor.
3.3.3.1
Calibrating a New Sensor
For a new sensor, continue the air calibration procedure as follows:
1.
Press the
key and then press the
Enter
key to select the
YES
menu option.
2.
Press the
Enter
key to acknowledge that you are resetting the
sensor lifetime clock
.
3.
As instructed, remove the sensor manifold to expose the new oxygen sensor to ambient air for about two
minutes. Then, press the
Enter
key to continue.
4.
A message indicating that the calibration is in progress will be displayed, and then the calibration data will be
shown. At that time, press the
Enter
key to save the calibration data and return to measurement mode.
Note:
A
second calibration of the new sensor should be performed within 1-2 days of the first calibration.
CAUTION!
The useful life of ppm sensors will be extended by minimizing exposure of the sensor to
air.
Summary of Contents for Panametrics oxy.IQ
Page 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Page 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Page 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Page 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Page 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Page 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Page 55: ......