3.5.4 Selecting mass flow units
1. Scroll to the desired
Mass Flow
units for the flow rate
display and press [
]. The available units for this prompt
are determined by the selection made at the
System
Units
screen. See
Table 5
for the available units.
Table 5: Available mass flow units
English
Metric
LB = Pounds
Kilograms
KLB = Thousands of LB
Metric Tons (1000 KG)
MMLB = Millions of LB
Tons (2000 LB)
2. Scroll to the desired
Mass Flow Time
units for the mass
flow rate display (from
seconds
to
days
) and press [
].
3. Scroll to the desired number of
Mdot Dec. Digits
(
digits
to the right of the decimal point in the mass flow rate
display) and press [
].
4. Scroll to the desired
Mass (Totalizer)
units for the
totalized mass flow rate display and press [
]. The
available units for this prompt are determined by the
selection made at the
System Units
prompt.
5. Scroll to the desired number of
Mass Dec. Digits
(
digits
to the right of the decimal point in the totalized mass
flow rate display) and press [
].
6. After completing the above steps, the
PanaFlow Z1G/
Z2G Process Gas Flowmeter
returns to the
Global
PROGRAM
window. Press [
] and scroll to
CH1
or
CH2
to
continue the initial setup programming.
3.6 Activating a channel
The
Channelx-ACTIV
submenu permits selection of the
desired measurement method. In addition, it is used
to activate/deactivate one or both of the channels in a
2-Channel
PanaFlow Z1G/Z2G Process Gas Flowmeter
.
To access the Channelx-ACTIV submenu, proceed as follows:
1. From the
Keypad Program
, scroll to
CH1
or
CH2
and
press [
].
2. In the
Channel PROGRAM
menu, scroll to
ACTIV
and
press [
].
3. Scroll to
Burst
to activate the channel/path, and press
[
].
Note:
Burst is automatically selected for a 1-Channel meter.
4. Scroll to one of the measurement methods described
below and press [
].
• Skan Only
is the preferred technique for locating the
acoustic signal and for high velocity measurements. It
is more robust in a noisy environment than the
Measure
technique.
• Skan/Measure
is the preferred technique to use for low
velocity measurements.
If
Skan Only
is selected at the above prompt, the meter
uses this technique exclusively. However, if
Skan/Measure
is selected, the meter uses
Skan Only
to find the acoustic
signal and then tries to use the
Skan/Measure
technique for
the actual measurement.
After completing the above step, the meter returns to the
Channel PROGRAM
menu. Proceed to the next section to
continue programming your meter.
3.7 Entering system data for a
channel
The
Channelx-System
submenu is used to enter system
parameters for the channel.
3.7.1 Accessing the channelx-system submenu
1. From the Channel PROGRAM menu, scroll to SYSTM and
press [
].
2. The first prompt asks for the
Channel Label
. Use the four
arrow keys to enter the desired label (in any numeric or
text combination up to five characters), and then press
[
].
3. The next prompt asks for the
Channel (Site) Message
.
Enter the desired text in the same manner as the
channel label with up to 15 characters, and then press
[
].
3.7.2 Selecting volumetric units
1. Scroll to the desired
Volumetric Units
for the flow rate
display and press [
]. the available units are listed in
Table 6
below.
2. Scroll to the desired
Volumetric Time
units for the flow
rate display (from
seconds
to
days
) and press [
].
3. Scroll to the desired number of
Vol Decimal Digits
(
digits
to the right of the decimal point in the volumetric flow
rate display) and press [
]
18
Summary of Contents for PanaFlow Z1G
Page 1: ...PanaFlow Z1G Z2G User s manual 910 321 Rev A...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 9: ...no content intended for this page 1...
Page 21: ...Figure 12 Remote mount electronics transducer and preamplifier wiring ref dwg 702 731 732 13...
Page 28: ...no content intended for this page 20...
Page 30: ...no content intended for this page 22...
Page 38: ...no content intended for this page 30...
Page 40: ...no content intended for this page 32...
Page 43: ...Table 13 Service record cont Date Description of service Performed by 35...