Copyright 2022 Baker Hughes Company.
English–PV 62X / PV 62X-IS Instruction Manual | 7
•
Make sure that there is no damage to the pressure station.
•
Make sure that there are no missing items.
Note:
Use only original parts supplied by the manufacturer.
Refer to Chapter
1 for a description of the pressure station controls.
5.2 Release the Pressure
To release all pressure in the system:
1. Open the refill valve counter-clockwise (1
turn).
2. Turn the volume adjuster fully counter-clockwise.
3. Slowly open the pressure release valve counter-clockwise (1
turn).
5.3 Filling and Priming
The following procedure describes how to fill the pressure station with hydraulic fluid and how to
prime the system.
If priming a long hose, keep the open end of the hose vertical to reduce trapped air.
5.3.1 Models without Priming Pump
1. Open the refill valve fully counter-clockwise.
2. Turn the volume adjuster fully clockwise.
3. Remove the pressure release valve. If the reservoir is full of the correct fluid, go to step
5.
4. Use the refill bottle to add more hydraulic fluid to the reservoir. Do not put too much fluid into
the reservoir. Leave a small air gap at the top of the reservoir.
Note:
To prevent contamination use only one type of hydraulic fluid in the pressure station.
5. Replace the pressure release valve. Turn fully clockwise until firmly shut.
3
1
2
5
4
1
7
6
2
3
8
Summary of Contents for Druck PV 62X