English–PACE Calibration Manual | 19

Appendix

 

B.  Pressure Units and Conversion Factors

To convert from pressure VALUE

 

1 in pressure UNITS

 

1, to pressure VALUE

 

2 in pressure 

UNITS

 

2, calculate as follows:

Pressure Units

Factor (hPa)

Pressure Units

Factor (hPa)

mbar

1.0

cmH

2

O @ 20°C

0.978903642

bar

1000.0

mH

2

O @ 20°C

97.8903642

Pa (N/m

2

)

0.01

kg/m

2

0.0980665

hPa

1.0

kg/cm

2

980.665

kPa

10.0

torr

1.333223684

MPa

10000.0

atm

1013.25

mmHg @ 0°C

1.333223874

psi

68.94757293

cmHg @ 0°C

13.33223874

lb/ft

2

0.4788025898

mHg @ 0°C

1333.223874

inH

2

O @ 4°C

2.4908891

inHg @ 0°C

33.86388640341

inH

2

O @ 20°C

2.486413

mmH

2

O @ 4°C

0.0980665

inH

2

O @ 60°F

2.487641558

cmH

2

O @ 4°C

0.980665

ftH

2

O @ 4°C

29.8906692

mH

2

O @ 4°C

98.0665

ftH

2

O @ 20°C

29.836983

mmH

2

O @ 20°C

0.097890364

ftH

2

O @ 60°F

29.8516987

VALUE 2

VALUE 1 FACTOR 1

FACTOR 2

-----------------------------

=

Summary of Contents for Druck PACE Series

Page 1: ...PACE Pressure Automated Calibration Equipment Calibration Manual...

Page 2: ...ator Connection for Input Sensor Calibration 10 5 5 PACE Indicator Connection for Barometric Sensor Calibration 10 6 Calibration Check 10 7 Calibration Adjustment 11 8 Calibration Procedure for CM3 8...

Page 3: ...Copyright 2010 Baker Hughes Company ii PACE Calibration Manual English...

Page 4: ...er warnings or cautions issued to protect the user and the equipment from injury or damage Use qualified technicians and good engineering practice for all procedures in this publication Pressure Do no...

Page 5: ...raction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemination of those substances in our...

Page 6: ...bbreviation Description ac Alternating Current ft Foot H2O Water Hg Mercury in Inch kg kilogram m Metre mbar millibar Pa Pascal PACE Pressure Automated Calibration Equipment psi Pounds per square inch...

Page 7: ...alibration Certificate shows the measurement uncertainty of the original pressure calibration standard To preserve uncertainty of the PACE calibration checks and adjustments must be performed using a...

Page 8: ...ight 2010 Baker Hughes Company English PACE Calibration Manual 5 Figure 2 PACE Control Module Reference Level Set the PACE units of pressure to one of the required units for calibration 1 Reference Le...

Page 9: ...equipment with the correct pressure rating Before applying pressure examine all fittings and equipment for damage Replace all damaged fittings and equipment Do not use any damaged fittings and equipme...

Page 10: ...8 Male to AN6 37 Male IO ADAPT BARB ISO 228 G1 8 Male to 1 4 Hose IO ADAPT G1 4 ISO 228 G1 8 Male to ISO 228 G1 4 Female IO ADAPT G1 8 ISO 228 G1 8 Male to ISO 228 G1 8 Female WARNING Parallel thread...

Page 11: ...bar 1450 psi see alternative sealing method in Figure 5 Figure 5 Alternative Sealing Method for 100 bar 1450 psi 1 PACE pressure connector 2 Bonded seal 3 ISO228 1 G1 8 pressure connector 4 NPT thread...

Page 12: ...port If a reference connection is unavailable fit snubber IO SNUBBER 1 to the PACE reference port 5 3 PACE Controller Connection for Barometric Sensor Calibration 1 Connect the output of the pressure...

Page 13: ...adds the barometric reading to a gauge range to produce a pseudo absolute range for CM2 and below For CM3 the PACE adds the barometric reading to an absolute range to produce a pseudo gauge range For...

Page 14: ...ecification is the sum of the Data Sheet precision and the long term stability specification with respect to the original pressure calibration standard It is recommended that an adjustment is performe...

Page 15: ...ity is difficult to maintain then the home screen showing pressure readings can be used with 7 digits If the number of digits is reduced then the error due to reduced resolution must be considered wit...

Page 16: ...full scale pressure 40 of full scale pressure 20 of full scale pressure 35 mbara 965 mbarg Atmospheric 0 mbarg 20 of full scale pressure 40 of full scale pressure 60 of full scale pressure 80 of full...

Page 17: ...Copyright 2010 Baker Hughes Company 14 PACE Calibration Manual English 60 of full scale pressure 80 of full scale pressure 100 of full scale pressure 0 mbarg...

Page 18: ...Copyright 2010 Baker Hughes Company English PACE Calibration Manual 15 PACE Controller Menus Appendix A PACE Menus and Screens A 1 PACE Controller Menus Calibration procedure next page...

Page 19: ...Copyright 2010 Baker Hughes Company 16 PACE Calibration Manual English Appendix A PACE Menus and Screens A 2 PACE Controller Screens...

Page 20: ...Copyright 2010 Baker Hughes Company English PACE Calibration Manual 17 PACE Indicator Menus A 3 PACE Indicator Menus Calibration procedure next page...

Page 21: ...Copyright 2010 Baker Hughes Company 18 PACE Calibration Manual English Appendix A PACE Menus and Screens A 4 PACE Indicator Screens...

Page 22: ...978903642 bar 1000 0 mH2O 20 C 97 8903642 Pa N m2 0 01 kg m2 0 0980665 hPa 1 0 kg cm2 980 665 kPa 10 0 torr 1 333223684 MPa 10000 0 atm 1013 25 mmHg 0 C 1 333223874 psi 68 94757293 cmHg 0 C 13 332238...

Page 23: ...Copyright 2010 Baker Hughes Company 20 PACE Calibration Manual English Appendix B Pressure Units and Conversion Factors...

Page 24: ...10 Baker Hughes Company This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are tra...

Reviews: