9.3 Disposal of Calibration Liquid
9.4 Disposal of Calibrator
See “Marks and Symbols on the Equipment” on page
ii for details of the Druck WEEE take-back
scheme.
10. Specification
10.1 Shared Specification
CAUTION
Dispose of the calibration liquid in accordance with the Technical
Safety Data Sheet.
Specification
Dry
TC and Liquid
TC Models
Calibrator:
Influence of the operating
temperature [0
°C to 50
°C (32
°F to
122
°F)] on the accuracy.
±
0.02
°C/°C (0.036
°F/°F)
Resolution reference temperature
setting range.
0.1
°C/F
Detection speed.
130
ms
Display:
Resolution
0.01
°C/F (-9.99 to 99.99), or 0.1
°C/F
Display units.
°C or °F (optional)
Two-line display:
- Reference temperature
- Target temperature
4-digit, 7-segment LED, 7
mm (0.28″) high
red = upper display, green = lower display
Display for sensor failure.
Sensor failure behavior.
The control is switched off.
Excess temperature behavior.
Thermal fuses interrupt the power supply on excessive temperature
inside the housing.
Temperature Control:
Controller outputs:
- Heater control
Solid state relay.
- Cooler control
Solid state relay.
- Fan control
Mechanical relay (lifetime = 100,000 switching cycles)
Environmental:
Ingress protection rating.
IP 20
Operating temperature range.
0°C to 50
°C (32°F to 122
°F)
Relative humidity.
30
% to 95
% RH (none condensing)
Transportation and storage
temperature.
-
10
°C to 60
°C (14
°F to 140
°F)