Copyright 2013 Baker Hughes Company.
26 | Dry
TC / Liquid
TC Temperature Calibrators–English
1. Press the
key when in calibration mode for approximately 5
seconds. The main menu
opens.
Figure
25: Control OFF Menu
appears on the upper display.
LED SET flashes on the lower display.
2. Press the
or
key until the desired operating mode is displayed.
3. Press the
key to confirm.
Note:
The controller is now switched on. The controller will enter into calibration mode. The
calibrator’s temperature will regulate to the desired set point temperature.
5.4.2 Manual Control
It is possible to switch off the automatic control of the calibrator and achieve the desired
temperature via manual control.
5.4.2.1 Switching On Manual Control
1. Press the
key for approximately 5
seconds. The main menu opens.
The last selected function appears on the upper display.
LED SET flashes on the lower display.
2. Press the
or
key until
(open loop) appears.
Figure
26: Manual Control (OPLO) Menu
appears on the upper display.
LED SET flashes on the lower display.