Copyright 2011 Baker Hughes Company.
40 | DPS 8000 Instruction Manual–English
Appendix
A.
CANopen Object Dictionary
The value is left adjusted with the remaining bits to the right side of the LSB set to zero.
OBJECT DESCRIPTION
INDEX
6100
Name
AI Input FV
Object Code
Array
Data Type
REAL32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Default Value
0.0
Sub-Index
001
Description
AI Input FV 1
Data Type
REAL32
Access
RO
PDO Mapping
Yes
Default Value
0.0
Sub-Index
002
Description
AI Input FV 2
Data Type
REAL32
Access
RO
PDO Mapping
Yes
Default Value
0.0
A.3.2 Object 6101h: AI Input Unit
OBJECT DESCRIPTION
INDEX
6101
Name
AI Input Unit
Object Code
Array
Data Type
UNSIGNED32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Value
Description
1
Pressure
2
Temperature
Summary of Contents for Druck DPS 8000 Series