Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 63
Settings
6.3 Settings
The RTD
-
INTERFACE function can be configured by tapping the Settings button which gives the
following:
Figure
6
-
5: RTD
-
Interface Settings
•
READING MODE
Allows the temperature measurement to be displayed as:
a. Temperature (°C or °F)
b. Resistance (Ohms)
•
UNITS
Allows change of units depending on the Reading Mode selected.
•
RTD PROFILES
Based on the Callendar-Van Dusen equation, the coefficients required for the RTD curve can
be selected using a set profile.
There are two standard default profile options provided which cannot be edited. These are the
“Default US” and the “Default EU”.