Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 45
Measure or Simulate a Resistance Temperature Detector (RTD)
c. Sine
•
AMPLITUDE - Select the Peak-to-Peak value.
•
OFFSET - Set an offset value (only applies when Sine and Triangle waveforms are
selected).
4.9 Measure or Simulate a Resistance Temperature Detector (RTD)
14 show CH1 set up to measure an RTD. A 4
-
wire
configuration gives the best accuracy; a 2
-
wire configuration has the lowest accuracy.
Figure
4
-
12: PT100 RTD Measure CH1 4-Wire (Range
-
200 to 850°C)
Figure
4
-
13: PT100 RTD Measure CH1 3-Wire (Range
-
200 to 850°C)