620G Instruction Manual | 27
Basic Calibrator Operation
3. Calibrator
3.1 Basic Calibrator Operation
Select CALIBRATOR from the Dashboard.
3.1.1 Layout
The calibrator screen displays measurement or source functions which are grouped in channels.
It is possible to display more than one channel on the calibrator screen. There is a total of six
separate channels listed as follows:
•
Electrical - Channels “CH1” and “CH2”.
•
Pressure (via PM
620 / PM
620T and MC
620G) - Channels “P1” and “P2”.
•
External Sensor (USB) - supports sensors such as TERPS, IDOS or RTD
-
INTERFACE.
•
Communications - supports HART®, FOUNDATION™ Fieldbus and Profibus®.
There are two display views in the CALIBRATOR screen when multiple channels are in use.
1 shows a normal view with three selected channels.
To expand the view on a particular channel, tap anywhere in the channel window area.
Figure
3
-
1: Calibrator Window – Normal View (Three Channels)