HART® SDC Application Preferences
1. Navigate to the calibration folder.
Figure
9
-
28: HART® Method – Analog Trim
2. Select “Trim analog” method.
3. Follow the on-screen instructions. If CH2 is set to Measure Current (24V), this can be used
to supply the reference meter value.
4. Read the CH2 value and enter this in the meter value text box using the keypad.
Figure
9
-
29: Enter Calibration Point
5. Select SET.
6. Repeat steps 3, 4 and 5 with “20
mA” selected. This will calibrate the transmitter’s output
current.
9.8 HART® SDC Application Preferences
Select the Preferences icon
to setup the HART® device search method.
The application allows searching by: