Copyright 2007 Baker Hughes Company.
62 | DPI 610/615 IS Instruction Manual–English
Appendix
A.
Datalog File Example
A.2 Typical Uploaded Procedure Data File (DPI
615)
The following gives details of a typical data file upload:
A.3 Record Type
The Record Type number contained in the header is a 16-bit field, with the individual bits
representing result formats as detailed below:
FILENAME:
TEST 6
DATE:
21/10/2006
TIME:
15:58
TRIGGER:
KEYPRESS
AMBIENT TEMP: 24.1 C
NO. OF POINTS 11
RECORD TYPE
0
CURRENT
mA PRESSURE INT bar
3.902 0.008 -0.65 %span PASS 21/10/2006 15:58:55
6.076 2.311 1.42 %span FAIL 21/10/2006 16:00:03
7.598 4.404 0.47 %span PASS 21/10/2006 16:00:57
8.085 5.023 0.41 %span PASS 21/10/2006 16:01:45
9.949 7.249 0.94 %span PASS 21/10/2006 16:02:42
13.002 11.300 -0.23 %span PASS 21/10/2006 16:03:30
17.005 16.102 0.77 %span PASS 21/10/2006 16:05:00
17.766 17.106 0.51 %span PASS 21/10/2006 16:06:07
17.812 17.105 0.80 %span PASS 21/10/2006 16:07:49
18.532 17.965 1.00 %span PASS 21/10/2006 16:16:35
20.007 20.212 -1.02 %span FAIL 21/10/2006 16:18:23
Table
A
-
1: Data File Record Types
Record Type
Description
1
P-Display
2
Switch Test
4
Leak Test
8
Input Channel: Min/Max ON
120
Combination of all Input Process Functions
896
Combination of all Output Process Functions
1024
mA Step Mode
2048
mA Value Mode
4096
mA Ramp Mode
8192
No Input Selected
16384
No Output Selected
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......