E.4 Selecting units for AI blocks
To select the units for the individual AI blocks:
1. Double click on the AI block for which you wish to set
the units (ANALOG_INPUT_1 or ANALOG_INPUT_2 in the
tree under GEFlow-XGF; see
).
2. Select the
scaling
tab and set the unit for the
measurement based on the flowmeter settings.
For example, if the flowmeter was set to use the metric
unit system and the PRIMARY_SELECTOR was set to use
velocity
you would choose
m/s
for the unit as shown in
Figure 61
below.
Figure 61: Units index drop down list
E.5 Resetting instrument totalizers
To reset the instrument totalizers:
1. Double click on the FLOW transducer block (in the
tree under GEFlow-XGF; see
2. Select the
others
tab and scroll down to the
CLEAR_TOTALIZERS listing.
3. Select
clear
from the drop down list box (see
Figure
62
below).
4. After the totals have been reset, select
normal
from the drop down list box to resume total
accumulation.
Figure 62: Clear totalizers drop down list