Operating Manual BaehrTec S2000
English– V101 – Status 07/2015
Page 29/52
For reasons of operational safety, we recommend that you do not turn off
the sounds!
You cannot make any changes to the menu item "Information"! This item is
for viewing purposes only.
If you exceed the service interval, negative values will appear for "Service
in", and the device will count the operating hours that have elapsed since
the last service date.
You can navigate through the options menu using the following buttons
•
Button Options Menu
(19)
opens the options menu
--> switches from menu level 2 to menu level 1 and back to the work mode.
•
Buttons for handpiece speed +/-
(14)
and
(15)
--> browse within menu level (cursor
)
•
Handpiece motor button
(13)
Select (Enter) / Access menu level 2
•
Button Home
(18)
Return to work mode
The magenta background indicates which menu item you are currently viewing.
The settings that have just been made will be indicated to you by the "x" on the right-hand
edge of the display.
4.6
Home Function (button (18))
This button has two functions:
By briefly pressing the Home button you will be taken directly from the options menu to the
Work mode (see section "Work mode").
By pressing this button for approximately 5 seconds, the base settings made by the factory
will be restored:
•
Speed of the handpiece motor = 6,000 rpm
•
Instrument rotation direction = clockwise
•
Language = German
•
Sound = Aqua
•
Background colour = white
•
Brightness = moonlight
•
Spray strength = normal
•
Spray pause = max. 3 seconds
The set-up of the base settings cannot be changed.
By pressing the button for approx. 5 seconds all setting are reset (also
the values of the memory buttons) to the factory settings.
The information (meter readings, etc.) will, of course, remain unaffected.
4.7
Memory buttons (buttons (30) - (32))
Your device has three memory buttons which you can use to save your preferred settings in
Work mode. These settings will be retained even if you turn off the device.
To now save the settings, please adjust them (instrument rotation speed, direction of
rotation and spray level) in work mode and keep the memory button you wish to save them
under pressed for longer than 2 seconds. Once the settings are saved, you will hear a
sound and the respective memory storage space is shown on the display.
All memory buttons are factory pre-set as follows: