Display correction factor
May 2019
LM_OG-AM_BA_02_1163
Page 9
9.2
Programming the datalogger
Below you will find the procedure to change time and date. The date can only be changed with the program button on the
backside, which is only accessible when the register has been removed from the meter. Removing the register from the meter
will break the sealing. This seal has to be restamped by the national authorities (MID).
The ERASE button removes all the logged dispenses; it does not affect time and date.
NOTE:
If no button has been pressed for more than 30 seconds, the register will automatically move back to the standard
display without saving the last entry.
10.
DISPLAY CORRECTION FACTOR
To verify the programmed correction factor, press
TOTAL
and
RESET
for 2 seconds.
Correction factor
Opens Alibi Memory
5s
5s
Opens Configuration
5s
5s
Move to DATE
register backside
Move to ERASE
2s
Change to YES
1x
Confirm ERASE
register backside
0.000
L
8793
TOTAL
19.09.12
0081
15.32
Zeit
20.09.10
Datum
Nein
Loschen
Ja
Loschen
----
Loschen
Level –
Date
19.09.2012 Current day
Last batch
Change the time by using the RESET and
TOTAL to move forward
Change the date by using the RESET
and TOTAL move forward
Use the RESET button to change to
YES
Use the program button to confirm
the ERASE
The ERASE will last for about 45s
Standard display
1.0000