Installation
Page 5/51
MID_M2000_BA_02_1807
B
A
200
0-
11
-M
ID
h
>
2 x
D
N
B
A
20
00-
12-
M
ID
> 3-5
x D
N
> 2
x
DN
BA2000-
1
3
-M
ID
B
A
2
0
00
-1
4
-MI
D
3.2.3 Meter location
Caution:
•
Do not install the detector on the suction sides of pumps. This could
damage the liner (in particular PTFE liners).
•
Verify that the pipeline is always filled on the measuring point, if not - a
correct or accurate measurement is not possible.
•
Do not install the detector on the highest point of a pipeline system. Gas
accumulation may follow.
•
Do not install the detector in downcomer pipes with free outlet.
•
Do not install the detector on pipes with vibrations. If pipes are strongly
vibrating, make sure that detector and amplifier are separated (separate
version).