Programming
Page 17/51
MID_M2000_BA_02_1807
5.2 Main menu
The following menu items are available to you in the main menu:
•
Meter setup
•
Measuring
•
Inputs and outputs
•
Reset of the totalizer
•
Setting of communication port
•
Specific counter settings
•
Counter information
•
Language selection
5.2.1 Meter setup
Scale Factor
You may chose this factor to optimize an accuracy or to achieve an
accuracy in flow measurement that is close or better than the
reproduceability of the meter. This factor corrects the actual flow rate in
percent (positively or negatively).
Note: Changing this value has an influence on the meter accuracy!
Empty Pipe
Detection
Fluid monitoring shows if measuring pipe has only partly been filled
with liquid. Monitoring can be switched on or off.
Note: On request, fluid monitoring can be adjusted to fluid’s conductivity
or to cable length.
Power Line
Frequency
For an optimum operation of the meter, set Power Line Frequency in
this menu at operating location.
Excitation
Frequency
This value shows in which frequency the meter’s coils are operated.
Supported frequencies are dependent on the configured power line
frequency and meter’s size.
Note: When selecting excitation frequency, make sure to always
observe that the ratio in respect of power frequency is integer.
50 Hz
60 Hz
1 Hz
1 Hz
3.125 Hz
3.75 Hz
6.25 Hz
7.5 Hz
12.5 Hz
15 Hz