TDV (TDVTNCED) GROUP
The Advanced group tab provides users a way to troubleshoot problematic applications, as well as setting up the flow meter
to best suit application needs . Pages within this tab should only be modified under the guidance of Badger Meter Technical
Support or with caution . Modifying values can alter flow measurement accuracy .
Meter
Display
Cal
System
Site
Log
I/O
Adv
Monitor
Transit
Signals
Waveform
Doppler
Figure 83: Advanced group
Signals Page
Figure 84: Advanced signals page
Monitor Signal
Select the signal type to be monitored from the
Monitor Signal
pull-down list and adjust the graph scale using the Axes
adjustments . This is used for reference only .
Monitor Signal
X axis
Description
Transit RX Signals
0…1000
Received Up/Down Signals
Transit RX Signals Analyzed
0…1000
Received Up/Down Signals, Analysis Portion
Transit RX Noise
0…1000
Received Up/Down Noise test signals
Transit XCorr
–40…40
Cross Correlations
Transit XCorr Alt
–40…40
Alternate Cross Correlations
Transit RX Envelope
0…1000
Received Up/Down signal Envelopes
Transit Rx Alt
0…1000
Alternate Received Up/Down Signals
Transit RX Waveout
0…1000
Transmitted Up/Down Waveforms
Pipe Thickness Signals
0…1000
Pipe Thickness Test Signals
Doppler Best Acorr
0…50
Doppler Autocorrelations
Doppler Best Rate
0…200
Doppler Auto Sample Rate Raw Data
Doppler Full Rate
0…200
Doppler 25 kHz Sample Rate Raw Data
Doppler HI Rate
0…200
Doppler 25/4 kHz Sample Rate Decimated
Doppler MED Rate
0…200
Doppler 25/16 kHz Sample Rate Decimated
Doppler LO Rate
0…200
Doppler 25/32 kHz Sample Rate Decimated
Table 19: Monitor signal parameters
Adv (Advanced) Group
Page 64
November 2016
HYB-UM-00090-EN-04