![Badger Meter Dynasonics DXN User Manual Download Page 58](http://html.mh-extra.com/html/badger-meter/dynasonics-dxn/dynasonics-dxn_user-manual_466112058.webp)
Logging Rate
(Once every #)
Logging Duration
Samples/Second
Minutes
Hours
Days
0 .1 second (10 Hz)
10
109
1 .8
0 .08
1 second
1
1092
18 .2
0 .76
2 seconds
0 .5
2183
36 .4
1 .52
5 seconds
0 .2
5458
91 .0
3 .79
10 seconds
0 .1
10,917
182
7 .58
20 seconds
0 .05
21,833
364
15 .2
30 seconds
0 .03333
32,750
546
22 .7
1 minute
0 .01667
65,500
1092
45 .5
2 minutes
0 .008333
131,000
2183
91 .0
5 minutes
0 .003333
327,500
5458
227
10 minutes
0 .001667
655,000
10,917
455
30 minutes
0 .00055556
1,965,000
32,750
1365
1 hour
0 .00027778
3,930,000
65,500
2729
2 hours
0 .00013889
7,860,000
131,000
5458
4 hours
6 .941
-5
15,720,000
262,000
10,917
12 hours
2 .315
-5
47,160,000
786,000
32,750
1 day
1 .157
-5
94,320,000
1,572,000
65,500
Table 17: Logging durations
Any log files gathered are stored with the site information .
Select Data Page
The
Log > Select Data
lets you choose the way the time is displayed for each logged entry and what kind of data is recorded in
the user fields .
Figure 75: Data selection (page 1)
Column 0E: Time
Column 0
can be set up for one of three time stamp formats .
• The conventional Year, Month, Day, Hour, Minute, and seconds .
• The Microsoft Excel® decimal time format .
• A simple counter that increments once for each logged entry recorded .
Column 1E: Flow
Column 1
always stores the current flow rate in the user-selected units .
Log (Data Logging) Group
Page 58
November 2016
HYB-UM-00090-EN-04