background image

CXX-UM-01850-EN-01

Setup Auto/Manifold Auto-ranging Trip Points

Use 

Setup Auto

 to set values as setpoints  Use setpoints to let the flow computer know when to change channels  See 

"Manifold System" on page 14

 for more information about channels 

1   Navigate to the 

SETUP AUTO

 option 

2   Press 

ALTEO

 to display the auto-ranging setpoints 

3   Press 

UP/DOWN 

arrows to

 

access each point 

4   Press 

CLEAO

 to clear the setting and enter a new number, or just enter a number to replace the old one 

5   Press 

ENTEO

 to save the new number  

Press 

ALTEO

 to back out of the menu and revert to the previous number   

Press 

CLEAO

 to delete any number you entered to start over 

Menu Option

Unit

Description

1up2

0 0 Hz Shows the trip point in Hertz to change from channel 1 (low flow) to channel 2 (medium flow)

2up3

0 0 Hz Shows the trip point in Hertz to change from channel 2 (medium flow) to channel 3 (high flow)

3down2

0 0 Hz Shows the trip point in Hertz to change from channel 3 (high flow) to channel 2 (medium flow)

2down1

0 0 Hz Shows the trip point in Hertz to change from channel 2 (medium flow) to channel 1 (low flow)

When setting these values, leave some leeway between setpoints  For example, if channel 1 is set and calibrated for  

20…800 Hz and channel 2 for 500…1200 Hz, set setpoint 1 to 2 at about 600 Hz and set setpoint 2 to 1 at about 550 Hz  This 

allows room for variance and avoids getting the flow computer into a switching loop 

Automatic Switching

When a setpoint is activated, the flow computer changes the selected channel  The flow computer then powers up a relay and 

closes a valve, or powers down a relay and opens a valve  In fail safe mode, the valves automatically open with no power and 

close with power 

Event

Oesult

Low flow signal from meter 1

Relays 1 and 2 powered up, closing valves for channels 2 and 3

Medium flow signal from meter 2

Relay 2 powered up, closing the valve for channel 3

High flow signal from meter 3

Both relays powered down

Setup Comms

Use this menu to configure communications with the PC 
1   Navigate to 

SETUP COMMS.

2   Press 

ALTEO

 to open the 

Setup Communications

 sub menu 

3   Press 

UP/DOWN arrows

 to cycle between sub menu options  Press 

ALTEO

 to change individual settings 

4   Press 

ENTEO

 to exit 

Option

Settings

Description

BAUD

9600, 1200 or 19200

Baud rate

READ

ONLY

The system is currently set to read only

ID

a, b, c and more

Select ID to scan from a…z lower case only

RS485

FAST, MEDIUM or SLOW

Delay for bus transceiver

COMMS

FAST, MEDIUM or SLOW

Delay between Char output

Programming Mode

Page 20 

February 2018

Summary of Contents for Cox 4050

Page 1: ...CXX UM 01850 EN 01 February 2018 User Manual Flow Computer Model 4050...

Page 2: ...CXX UM 01850 EN 01 Flow Computer Model 4050 Page ii February 2018...

Page 3: ...Using the Remote Control or Front Panel Keypad 14 Manifold System 14 Menu Flowchart 15 Navigating the Menus 16 Primary Menu 16 Secondary Menu 16 Changing the Menu Option Names and Unit Labels 18 Progr...

Page 4: ...l Address 24 Computer Interface Information 25 Pin Settings 25 Code Example 25 Special Instructions 27 Calibrating the PT100 27 Description of Unit Calculations 27 K Factor Calculation 27 Hz u K facto...

Page 5: ...quest an inspection by the carrier s agent within 48 hours of delivery and file a claim with the carrier A claim for equipment damage in transit is the sole responsibility of the purchaser SAFETY CONS...

Page 6: ...lassification see manufacturer s nameplate If the operator detects errors or dangers or disagrees with the safety precautions taken inform the owner or the principal responsible party Adhere to the lo...

Page 7: ...to remote facilities such as alarms indicators chart recorders Programmable Logic Controllers PLCs and more 0 10V or 4 20 mA Configurable analog output reference parameter Pulse Output Card Three col...

Page 8: ...Terminal Name RELAYS ANALOG IN FREQUENCY Slot 3 Slot 2 Slot 1 1 NO RELAY 1 SIG 1V OR I SPECIAL IP 2 C RELAY 1 0V SPECIAL IP 3 NO RELAY 2 SIG 2V OR I EXITATION VOLTAGE 4 C RELAY 2 0V SIGNAL 1 5 NO RELA...

Page 9: ...3 SW1 input 4 SW2 input 5 SW3 input 6 0V Total Inhibit Input By shorting SW1 terminals 3 and 6 on the Hub Connector the totalization on all inputs are inhibited from counting up Manifold High flow Met...

Page 10: ...he front panel keypad Front Panel Keypad Figure 3 Front panel keypad Key Key Name Key Function Up Arrow Down Arrow Scrolls through information screens Left Arrow Right Arrow Toggles between Primary an...

Page 11: ...ow Scrolls through information screens Left Arrow Right Arrow Toggles between Primary and Secondary menus ALTER Selects option to be changed SHIFT Prepares the remote to execute a secondary function f...

Page 12: ...TCH Used to enter 1 Sub function activates batching function 2 F1 Used to enter 2 Sub function activates function 1 3 F2 Used to enter 3 Sub function activates function 2 4 F3 Used to enter 4 Sub func...

Page 13: ...ntrol function 1 Press LEFT RIGHT arrows on the front panel at the same time to enter the Front Panel Test mode 2 Press UP DOWN arrows to highlight ENABLE IR 3 Press ALTER to select N for ENABLE IR On...

Page 14: ...e one in Figure 5 See the Turbine Meter Manifold System data sheet available in the Resource Library at www badgermeter com for more information The flow computer receives signals from the meter measu...

Page 15: ...rq Constant Setup Comms BAUD Read Only ID RS485 Comms Sort Curve 0 Yes Analog Inputs Pt100 Factor Set Viscosity V01 0 0 C V01 0 0 CTS V02 0 0 C V02 0 0 CTS Set SG SG01 0 0 C SG01 0 0 SG02 0 0 C SG02 0...

Page 16: ...Flow Rate for channel 1 2 or 3 L hr m1 m2 or m3 Mass Flow of turbine 1 2 or 3 kg hr v1 v2 or v3 Total Volume of turbine 1 2 or 3 L w1 w2 or w3 Total Weight of turbine 1 2 or 3 kg hr Temp Temperature f...

Page 17: ...e Channel to change which channel is active This option displays the current channel number Press ALTER to change the channel number NOTE N To use this option make sure Manifold is set to OFF otherwis...

Page 18: ...o flash at the far right side of the screen 3 Change the unit labels a Press LEFT RIGHT arrows to change the cursor position b Press UP DOWN arrows to select the character for that position 4 Change m...

Page 19: ...electing the secondary menu wait 20 seconds and the primary menu displays Automatic Lockout In some cases the flow computer itself inhibits the ALTER key as it may be performing a sequence of events t...

Page 20: ...e if channel 1 is set and calibrated for 20 800 Hz and channel 2 for 500 1200 Hz set setpoint 1 to 2 at about 600 Hz and set setpoint 2 to 1 at about 550 Hz This allows room for variance and avoids ge...

Page 21: ...ntry 5 Press ENTER to set the data The screen displays the entered data with the curve and point number 6 Press ENTER to exit this menu and return to Programming Mode Ch Setup Use this menu option to...

Page 22: ...ts menu to setup the PT100 input for calibration 1 Navigate to ANALOG INPUTS 2 Press ALTER to enter the sub menu The screen displays PT100 factor 3 Press ALTER or ENTER to see the PT100 Factor Alterna...

Page 23: ...Fitted This option is for information only 1 Press ALTER to display the first card slot 2 Press UP DOWN arrows to display each slot and its status Example of Cards Fitted Option Example Description 1...

Page 24: ...memory address space used to store constant and curve data The Hub is the main processor the card connected to the serial link while the SLOTs need to be addressed as follows Description ID View Hub S...

Page 25: ...2 If eb 00 Then GoTo the_res_is_zero sethex Right result 6 e Val h eb 127 bin_data For p 1 To 6 Select Case UCase Mid sethex p 1 Case 0 nibble 0000 Case 1 nibble 0001 Case 2 nibble 0010 Case 3 nibble...

Page 26: ...str Format dec 0 0 Case 2 result_str Format dec 0 00 Case 3 result_str Format dec 0 000 Case 4 result_str Format dec 0 0000 Case 5 result_str Format dec 0 00000 Case 6 result_str Format dec 0 000000 C...

Page 27: ...t be programmed for correct operation of the computer NOTE N Data for each curve may be entered in any order The computer sorts all data into numerical order when you return to the main menu on comple...

Page 28: ...rature versus S G curve 8 Totalizes the mass flow rate NOTE N You can prevent totalization by shorting together terminals 34 and 36 at the rear terminal block Auto ranging Mode The computer operates i...

Page 29: ...CXX UM 01850 EN 01 Hz u K factor system Figure 8 Hz u K factor system Special Instructions Page 29 February 2018...

Page 30: ...full scale least significant digit Pulse Outputs TTL or open collector outputs with a range of 1 Hz 1 kHz Relays 0 25A at 240V AC 9 dry relay Power Supply Mains input or DC input 110 240V AC is select...

Page 31: ...CXX UM 01850 EN 01 User Manual INTENTIONAL BLANK PAGE Page 31 February 2018...

Page 32: ...uarter Building Wing C Office C209 Dubai UAE 971 4 371 2503 Slovakia Badger Meter Slovakia s r o Racianska 109 B 831 02 Bratislava Slovakia 421 2 44 63 83 01 Asia Pacific Badger Meter 80 Marine Parade...

Reviews: