Badger Meter Blancett B2800XP User Manual Download Page 11

Password

The password setting restricts access to 

PROGRAM

 mode  Initially, the password is set to all zeros and any user can modify 

the paramter settings  To change the password, press 

ENTEO

 at the 

PASSWORD

 prompt  The first digit flashes  Press 

UP

 to 

increment the digit and 

OIGHT

 to advance to the next digit  After entering all digits, press 

ENTEO

 to store the password and 

advance to 

RST PSWD

  The new password is now required to enter 

PROGRAM

 mode  With this password set, any user is able to 

reset the stored totals on the monitor 

OTEE:

N

This password will allow the operator to manually reset totals 

Oeset Password

The reset password parameter restricts resetting the totals on the monitor  The 

PASSWORD

 must also be set to restrict the total 

reset  Initially, the password is set to all zeros and any user can reset the stored totals on the monitor  To change the password, 

press 

ENTEO

 at the 

RST PSWD

 prompt  The first digit flashes  Press 

UP

 to increment the digit and 

OIGHT

 to advance to the next 

digit  After entering all digits, press 

ENTEO

 to store the password and return to the 

METER

 parameter  The reset pasword is 

now required to reset the totals on the monitor 

OTEE:

N

Entering a password in the 

PASSWORD

 screen and leaving the default password in the 

RST PSWD

 screen allows for 

total resets (not requiring a password), but restricts programming modification 

Oeturn to Oun Mode

After entering all parameters, press 

MENU 

to return to 

RUN

 mode  

Oun Mode

Oeset Total

To reset the monitor’s total display, in 

RUN

 mode, press 

MENU

 and 

ENTEO

 simultaneously until 

TOTAL RST

 starts to flash  The 

display returns to the 

RUN

 mode at the conclusion of the reset procedure 

Store Total

The current total can be manually stored in the monitor’s flash memory  This procedure may be needed prior to changing the 

settings or replacing the battery  Press and hold 

ENTEO

 for two seconds  The display flashes 

TOTALSVD

 and then returns to 

RUN

 mode 

Automatic Store Total

The monitor is equipped with a store total feature that works automatically, saving the current total to flash memory  The 

frequency of saves depends on the power supply option chosen  

 

Battery Powered: Once per hour and just before a low battery condition turns the unit off  

 

Loop Powered:  Once every ten minutes 

Programming

Page 11 

March 2017 

DSY-PM-00279-EN-03

Summary of Contents for Blancett B2800XP

Page 1: ...Flow Monitor B2800XP Explosion Proof Flow Monitor Advanced DSY PM 00279 EN 03 March 2017 User Manual...

Page 2: ...Mode 11 Additonal Scaling Parameters 12 Flow 4 mA Setting 12 Flow 20 mA Setting 12 4 20 mA Calibration 12 4 20 mA Test 12 Additional Input Options 13 Battery Replacement 13 Maintenance 14 Additional...

Page 3: ...serious personal injury Indicates a hazardous situation which if not avoided could result in severe personal injury or death Indicates a hazardous situation which if not avoided is estimated to be ca...

Page 4: ...lectable units viewing in run mode and prompts for variables in program mode The display can show rate total or alternate between both rate and total MENU ENTER Programming Buttons Figure 1 B2800XP fl...

Page 5: ...amage to the cover threads and glass lens when glass lens cover is used 5 Pull wires into enclosure making certain they are long enough to make the required connections and to remove the instrument or...

Page 6: ...b Size Kit Number Meter Sizes 1 2 in NPT Hub B280 742 3 8 in 1 2 in 3 4 in with 1 2 in NPT End Fittings 1 in NPT Hub B280 737 All Sizes with 1 in NPT End Fittings and Larger Installation Kit P N B280...

Page 7: ...ion turn on the lock out feature to prevent unauthorized access or changing the meter s setup parameters MENU ENTER Figure 3 Front panel Buttons MENU Enters and exits PROGRAM mode UP Scrolls forward t...

Page 8: ...If the current setting is correct press ENTER to advance to the next parameter To change the current setting press UP or RIGHT to scroll to the bore size of your meter Press ENTER to save and advance...

Page 9: ...rams were chosen as the RATE UNT or TOTL UNT the monitor advances to the SP GRAV parameter Specific Gravity At the SP GRAV parameter press ENTER The first digit of the current specific gravity flashes...

Page 10: ...plitude of the pulse is dependent on the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC Linearization Linearization enhances the accuracy of the display T...

Page 11: ...password and return to the METER parameter The reset pasword is now required to reset the totals on the monitor NOTE N Entering a password in the PASSWORD screen and leaving the default password in t...

Page 12: ...4 20 mA Calibration The 4 20 mA Calibration parameter is the fine adjustment of the 4 20 mA output The 4 mA setting is typically between 35 50 To set the 4 mA value connect an ammeter in series with...

Page 13: ...Power Figure 7 Circuit board layouts BATTERY REPLACEMENT Battery powered monitors use a single 1 5V D size alkaline battery When replacement is necessary use a clean fresh battery for continued troub...

Page 14: ...ates some form of alternating signal that can be raw that is no format amplification or wave shaping can be done prior to transmission to the readout The output pulse rate is related proportionally to...

Page 15: ...t 2 25 Pulse Output Signals for Non linear Processes Few flow meters actually behave in a linear way There is always some uncertainty about the exact flow at a given reported input value For many comm...

Page 16: ...nitor pig tail This should stop the noise If not then try reorientating the meter to a new location Flow rate indicator bounces This usually indicates a weak signal Replace pickup and or check all con...

Page 17: ...ity 30 mV p p Over Voltage Protected 30V DC Frequency Measurement Accuracy 0 1 Temperature Drift 50 ppm C maximum Optional Analog Output Loop Powered Version Type 4 20 mA current loop Resolution 1 400...

Page 18: ...RT NUMBER INFORMATION NOTE Advanced B2800 monitors only The default is gallons per minute Program Level A Advanced B 28 A X X X X Power Option B Battery L Loop Mounting Style X Explosion proof Units o...

Page 19: ...INTENTIONAL BLANK PAGE Page 19 March 2017 DSY PM 00279 EN 03...

Page 20: ...Colonia Guadalupe Inn CP 01050 M xico DF M xico 52 55 5662 0882 Europe Eastern Europe Branch Office for Poland Latvia Lithuania Estonia Ukraine Belarus Badger Meter Europe ul Korfantego 6 44 193 Knur...

Reviews: