8.5 MISC
In the
Miscellaneous
submenu the limits and delay settings can be changed
Inlet Tubing volume:
Setting to include the volume of the inlet hose in the
measurement process
Show custom unit:
Selecting a check mark
will show a Custom
calculated value instead of pmol/min.
Delay after Cycles:
Enter a value will stop the measurement process for a
period of time after the entered cycles (time cam be selected by Delay time)
Delay Time:
This is the amount of delay time before a new measurement
cycle will start. Applicable only when the option; ” Delay after Cycles” is
activated
Detection limit minimum:
A detection limit above this value can be seen as
unreliable. An Erroneous Measurement result will occur.
Detection limit threshold:
Internal calculation value (100 %)
10 ml chamber volumes:
Selecting a check mark
will activate settings
for a 10 ml reaction chamber. Deselecting a check
mark will activate
settings for a larger reactor.
Custom unit:
For setting a calculated Custom value and unit.
Bubble sensor sensivity:
This is the sensitivity of bubble check for incoming
water and may occur an alarm. (value can be set lower when incoming water
contains air bubbles).
2nd inlet:
When selecting a check mark here,
all pump steps, except
the filtration step ,are done via 2
nd
inlet. (this will also add valve 3.3 in service
menu
Enclosure Heating:
When selecting a check mark here
will turn on the
cabinet heater (Optional)
Enclosure setpoint:
Setpoint for cabinet temperature.
Post filtration from 2nd inlet
When selecting a check mark here
This
inlet will be used for rinsing the reactor with external fresh- or demi water to
optimize a measurement.
Post filtration volume:
Setting volume to use for post filtration
Summary of Contents for microLAN
Page 1: ...Software Manual Version V0 1 15 Draft Version Date 26 02 2020...
Page 2: ...Page 2 of 28 BACTcontrol Software Manual V0 1 15 This page is intentionally left blank...
Page 25: ...microLAN b v BACTcontrol Software Manual V1 15 Whit Contact details can be added...
Page 28: ...Page 28 of 28 BACTcontrol Software Manual V0 1 15 Figure 21 Advanced screen...