background image

Tru Pointe

®

 Ultra HD

Ultrasonic Leak Detector

Instruction 0028-9001

Operation and Maintenance

Rev. 1 - May 2010

The Measurable Difference

®

Summary of Contents for Tru Pointe Ultra HD

Page 1: ...Tru Pointe Ultra HD Ultrasonic Leak Detector Instruction 0028 9001 Operation and Maintenance Rev 1 May 2010 The Measurable Difference...

Page 2: ...warranty of Seller Bacharach warrants that this product conforms to the Product Description contained in this literature Bacharach makes no other warranty whether expressed or implied including warra...

Page 3: ...8011 2 Features 2 3 Specifications 3 4 Operation 4 Physical 4 Taking Readings 4 General 4 Airborne Leak Detection 5 Operating Tips 6 Electrical Arcing 6 Electrical Corona 6 Using the Touch Probe 7 The...

Page 4: ...NOTES...

Page 5: ...G and the warning symbol at left in this manual denotes a potential hazard associ ated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not c...

Page 6: ...to detect leaks at the speed of sound without the delay seen with gas dependent sniffers What s in the Kit 0028 8001 Tru Pointe Ultra HD Kit Tru Pointe Ultra 0028 7000 Instruction Manual 0028 9001 Ca...

Page 7: ...Connector 3 5 mm Stereo Weight 0 3 lb 138 g including battery Electrical Sensitivity 80 db V bar Ultrasound Converter Type Heterodyne Frequency Response 34kHz to 42kHz Heterodyne Oscillator Analog He...

Page 8: ...he detector 2 Slide the Sensitivity Adjust toward the top of the instrument until it stops The Tru Pointe Ultra HD is now at full sensitivity NOTE Always begin testing at full sensitivity and adjust a...

Page 9: ...reduce the sensitivity until the LED s disappear 2 While pressing the On Off Button scan the area of the suspected leak in a side to side motion 3 While scanning listen to differences in the sounds d...

Page 10: ...electronic devices Place your body between the source of the sound and the leak point Electrical Arcing Electrical arcing is an electric current flow in air and can be a very destructive and dangerous...

Page 11: ...from the Tru Pointe Ultra HD 2 Screw the Touch Probe onto the sensor 3 The Tru Pointe Ultra HD is now ready to detect solid borne sound Thermal Expansion Valves TXV The following steps can be used to...

Page 12: ...y the SoundBlaster using the same method as detecting a standard airborne leak NOTE Using the SoundBlaster inside a symmetrical object i e a cylindrical tank can cause standing waves which can result...

Page 13: ...aks with the Tru Pointe Ultra around the perimeters of the doors and windows 3 If the Tru Pointe Ultra HD reading is at 10 reduce the sensitivity until the detector reads 4 5 and begin scanning again...

Page 14: ...e the batteries use only high quality 9 volt alkaline for both the Tru Pointe Ultra HD and the SoundBlaster and use the following instructions 1 Hold the unit with the Battery Door facing up 2 Press l...

Page 15: ...11 Tru Pointe Ultra Instruction 0028 9000 NOTES...

Page 16: ...The Measurable Difference Bacharach Inc Phone 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 www mybacharach com help mybacharach com...