Appendix B - Specifications
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
5-2
5.2
Appendix B - Specifications
MEASUREMENT:
0
– 1 / 0 – 10 / 0 – 25 / 0 – 50 % Oxygen – Fixed or Auto Range
DISPLAY:
0.75” 7-segment LED digital display, 4 characters
Displays Oxygen from 0.0 to 50.0 Percent.
Resolution:
0
– 9.9 %: X.X
0
– 50.0 %: XX.X
Color -
Coded LED’s for system status:
NORMAL:
Green
FAULT:
Yellow
LOW FLOW:
Red
O
2
ALARM:
Red
SIGNAL
INTERFACE:
Serial Service Port:
RS-232
Analog Voltage Output:
0
– 1, 0 –5, 0 – 10V
Analog Current Output:
0
– 20mA
4
– 20 mA
Range ID Voltage:
0
– 1 % O
2
: 5.63 V
0
– 10 % O
2
: 6.25 V
0
– 25 % O
2
: 6.88 V
0
– 50 % O
2
: 7.50 V
RELAY OUTPUTS
O
2
Alarm:
Field Adjustable Form C (DPDT) Voltage-free
5A@250Vac, 5A@30Vdc. Configurable to fail-
safe/non-fail-safe
and
ascending/descending
activation
ICS:
Field Adjustable Form C (DPDT) Voltage-free
5A@250Vac, 5A@30Vdc. Configurable to fail-
safe/non-fail-safe
and
ascending/descending
activation
Calibration
and Fault:
Non-adjustable Form B (SPST) Voltage-free
5A@250Vac, 5A@30Vdc, non-Fail-Safe activation,
non-configurable.
AVAILABLE
SENSOR TYPES
Model GP Sensor
Model CAG-250 Nylon Sensor
Model IT Sensor
RESPONSE TIME
T
90
< 20
ACCURACY
+ 2.0% FSD @ STP