System Setup
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
4-13
4.1.2.10.14
Test with Percentage Full Scale
This menu may be used to test the Analog Current and Voltage output in terms of percentage full
scale. Any percentage between 0.0 and 100.0 may be entered at the prompt. Navigate from
Analog Output Setup menu to the Analog Output Calibration
menu by typing “3” on the RS-232
terminal. Then, type “T” to access the Test with Percentage Full Scale menu.
4.1.2.10.15
Force Manual Voltage Output Range to 0 to 25%
This setting is used to reset the voltage output range to the default values, which are between 0 to
25% oxygen. Entering “Y” at the prompt resets the Analog Voltage zero to the default. Entering
“N” at prompt retains the current setting. This menu is accessed from the Analog Output Setup
menu by typing “5” on the RS-232 terminal
4.1.2.10.16
Force Manual Current Output Range to 0 to 25%
Thes menu forces the Manual Current Output Range to the default values, which are between 0 to
25%. Entering “Y” at the prompt resets the Analog current zero and full scale. Entering “N” at
prompt retains the current setting. This menu is accessed from the Analog Output Setup menu by
typing “6” on the RS-232 terminal
4.1.2.11
(U04) Display/Analog Output Range
The RS-232 Display/Analog Output Range menu U04 provides access to the Analog Output
Range (Current and Voltage) settings. The Analog Output Range may be set to 0 (0-1% Fixed), 1
(0-10% Fixed), 2 (0-25% Fixed), 3 (0-50% Fixed), 4 (0-1/10/25/50% Auto-range) or 5(Manual-
Range). This menu is accessed from the Setup Main Menu by typing “4” on the RS-232 terminal.
To navigate backwards
, use the <Esc> or “Q” key on the RS-232 terminal.
4.1.2.12
(U50) RS-232 Serial Setup Menu
This menu provides access to set the RS-232 serial communications options. It is accessed from
the Setup Main Menu by typing “5” on the RS-232 terminal. To navigate backwards, use the
<Esc> or “Q” key on the RS-232 terminal.
4.1.2.12.1 Baud Rate
This menu sets the RS-232 baud rate. The baud rate can be set to 0 (110bps), 1 (300bps), 2
(1200bps), 3 (2400bps), 4 (4800bps), 5 (9600bps), 6 (19200bps) or 7 (38400bps). This menu is
accessed from the RS-
232 Serial Setup menu by typing “1” on the RS-232 terminal.