P/N: 3015-3843 (Rev 2) 

BACHARACH

 

1

 

 

Testing HGM 300 with Gas Verification Kit 

Using PC Software  

 

3015-3843  Rev. 2 

February 2017

 

 

 

This procedure describes the use of gas verification kits with the HGM PC software. 

 

MATERIALS  REQUIRED 

 

Reference gas 

 

Variable flow regulator 

 

Approximately 5 feet of 3/16” tubing 

 

Plastic stem connector 

 

Computer with HGM 300 software installed 
(available at 

www.mybacharach.com

) 

 

PROCEDURE 

1.

 

Power up the HGM. The green LED on the monitor will blink until the warm-up cycle is complete. 

2.

 

Connect the variable regulator to the reference cylinder and connect the tubing to the variable regulator.   
Connect the plastic stem connector to the other side of the tubing.  

3.

 

Connect the RS232 cable from the PC to the HGM’s RS232 connector located in lower left side of the 
monitor.   

NOTE:  

For instructions on PC software and setup, see the instruction manual. 

4.

 

Launch the PC software. A series of beeps will sound as the software reads all the HGM registers.  

5.

 

Verify Rezero Mode is set to auto by reading the system parameter list. If it is not set to Auto, do the 
following:  

a.

 

Select the “Edit Pull down Menu”. 

b.

 

Choose the “System” item from the selection list.  

c.

 

Change the Rezero Mode to “Auto”. 

d.

 

Press the Enter key to update the setting.  

 

The program should respond with “System Data Updated”.  

 

(On version 1.0 software it is necessary to select the “HGM Pull down Menu” and then send the 
setup in order to confirm the changes) 

6.

 

Select the “HGM Pull down Menu” and choose the setup item from the menu list. This will update the setup 
from the HGM. 

7.

 

Select the “Edit Menu” and choose “Zones” from the menu item list. 

8.

 

Select zone by pressing the Enter key. Modify the parameters for Zone 1: 

a.

 

Change the refrigerant selection to the reference cylinder type 

b.

 

Change the distance to 5 ft. 

c.

 

Press the Enter key, then the ESC key. 

d.

 

Verify that the PC software recognizes that the HGM Zone data was updated.  

NOTE: Be sure to wait for response from the PC software. 

Selecting another menu before the response is 

echoed will cause the communications to timeout.  On version 1.0 software it is necessary to select the “HGM 
pull down menu” and then send the item to the setup in order to confirm the changes. 

9.

 

Select “HGM Menu” and choose “Zone Hold” from the menu list.  