40
BTEK T403S Indicator User Instructions
Menus
4.7 Audit menu
The Audit menu is shown in
.
Figure 4.4 Audit menu
Use this menu to display audit counters for configuration and calibration and to print the
information. Each is explained below:
4.7.1 Counter
Audit
Counter
1.
Access the Audit menu and press
SELECT
…
countEr
is displayed. This has two counters that tell you how many
times the indicator has been configured and calibrated.
2.
Press
SELECT
…
conFig
is displayed.
3.
Press
SELECT
again …
A number appears showing how many times the indicator has been
configured.
4.
Press
ZERO
…
conFig
is displayed.
5.
Press
UNITS
…
cALib
is displayed.
6.
Press
SELECT
…
A number appears showing how many times the indicator has been
calibrated.
7.
Press
ZERO
…
cALib
is displayed.
8.
Press
TARE
…
countEr
is displayed.
Audit
Counter
Config
Calib
Port 1
Port 2
USB
SELECT
=
TARE
=
=
UNITS
=
ZERO
=
F1
= Escape or
Displays
number of
configurations
Displays
number of
calibrations
Summary of Contents for T403S