13
3.2 Castors
• The device is equipped with 4 wheels. Two of these wheels are equipped with brakes. Once it is in its final
position, the brakes of the front wheels must be locked to immobilise it and stop it from rolling away .
3.3 Fixation
• During transport in a vehicle, the equipment must be secured with an appropriate belt system (the device
is standardly delivered with an adequate belt system). If other belts are used, they must fulfill the
requirements of the EN standard 12195-2. Each single belt must be able to transmit a traction power of at
least 1500 daN .
WARNING
• Not activating the breaks, may cause the unit to roll away and cause damage to the unit and physical injury.
• Not securing the device with the appropriate belt system, may cause damage to the unit during transit and
physical injury.
3.4 Initial cleaning and disinfection
CAUTION
• The use of hydrogen peroxide H2O2 (VHP) for the disinfection (sterilization) of the device is not suitable.
• Before using the device, it should be cleaned inside and outside.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scouring
powder, steel wool, abrasive sponges or chemical solvents.
• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer,
provided they are recommended by the national organizations. For disinfecting small areas, we recommend
using a concentrated alcoholic agent.
• Other than the cleaning procedures described in this manual, follow the internal policy and procedures on
cleaning, disinfection and maintenance established by your facility.
• Refer to section 8 for more information on cleaning.
3.5 Connection of remote alarm
• Connect the remote alarm located on the back of the unit to your surveillance system by respecting the
connecting diagram in section 6.7.
Summary of Contents for MT100
Page 1: ...820 9506 71 ed0821 INSTRUCTIONS FOR USE MT100 BLOOD REFRIGERATOR INCUBATOR...
Page 2: ...2...
Page 5: ...5 1 Model Overview Group Model name Blood Refrigerator Incubator MT100...
Page 24: ...24 12 Spare parts Container door and inner equipment...
Page 25: ...25 Electronic and cooling unit...
Page 27: ...27 13 Wiring diagram 294 8062 37...
Page 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...