12
3 Installation Procedure
WARNING
• All installation work and adjustments to the refrigerator must be carried out by qualified personnel. Work
performed by persons with insufficient technical knowledge may adversely affect the performance or
cause physical injury or damage to the equipment.
• Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved.
Deterioration of the insulation may cause current leakage or electric shock.
• When transporting or moving the device, be aware of its weight. Do not try to move a fully loaded device.
3.1 Location
• The following section provides information on how to install and place the device in your location. Please
note that all indicated distances are minimum distances.
WARNING
• Do not use the device outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
• Ensure that the unit is steady and perfectly straight and does not knock against anything next to it. If
the floor is not strong enough or the installation site is not adequate, this may result in injury from the
unit falling or tipping over.
• Never install the unit where acid or corrosive gases are present as current leakage or electric shock may
result due to corrosion.
• The wall socket must be easily accessible.
• When the unit is in its final position it is important to immobilise the unit using the feet.
CAUTION
• Optimal function of the appliance is given at an ambient temperature indicated in the technical data (see
chapter 10) with a maximum relative humidity of 95%.
• Ensure that there is sufficient room around the unit for air circulation. Refer to images 1 and 2 for
minimum distances to a wall or adjacent units. A good air circulation especially around the compressor
is essential for trouble-free operation. Keep the ventilation openings of the compressor cover always free
from any obstructions. Make sure that a minimum distance is provided between the unit and any wall or
other device located next to it.
• The device must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it close to a
heat source. The area must be protected against rain and dust.
• Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
• This device cannot be used/is not for use in facilities at altitudes of 2000m or higher.
• Two spacers are included with the MRB3000SDD. These must be attached to the rear of the unit as shown
in the figure, before positioning the unit against the wall.
min.
10 cm
min.
3 cm
MB 3000 G
spacer
image 1: minimum distance to wall or other devices
image 2: spacer installation