3   Installation Procedure

WARNING

• All installation work and adjustments to the refrigerator must be carried out by qualified personnel. Work

performed by persons with insufficient technical knowledge may adversely affect the performance or
cause physical injury or damage to the equipment.

• Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved.

Deterioration of the insulation may cause current leakage or electric shock.

• When transporting or moving the device, be aware of its weight. Do not try to move a fully loaded device.

3.1  Location

• The following section provides information on how to install and place the device in your location. Please

note that all indicated distances are minimum distances. 

WARNING

• Do not use the device outdoors. Current leakage or electric shock may result if the unit is exposed to

rain water. 

• Ensure that the unit is steady and perfectly straight and does not knock against anything next to it. If

the floor is not strong enough or the installation site is not adequate, this may result in injury from the
unit falling or tipping over.

• Never install the unit where acid or corrosive gases are present as current leakage or electric shock may

result due to corrosion.

• The wall socket must be easily accessible.
• When the unit is in its final position it is important to immobilise the unit using the feet.

CAUTION

• Optimal function of the appliance is given at an ambient temperature indicated in the technical data (see

chapter 10) with a maximum relative humidity of 95%.

• Ensure that there is sufficient room around the unit for air circulation. Refer to images 1 and 2 for

minimum distances to a wall or adjacent units. A good air circulation especially around the compressor
is essential for trouble-free operation. Keep the ventilation openings of the compressor cover always free
from any obstructions. Make sure that a minimum distance is provided between the unit and any wall or
other device located next to it. 

• The device must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it close to a

heat source. The area must be protected against rain and dust.

• Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
• This device cannot be used/is not for use in facilities at altitudes of 2000m or higher. 

•    Two spacers are included with the MRB3000SDD. These must be attached to the rear of the unit as shown

in the figure, before positioning the unit against the wall.

min.

10 cm

min.

3 cm

MB 3000 G

spacer

image 1: minimum distance to wall or other devices

image 2: spacer installation

Summary of Contents for MRB 3000 SDD

Page 1: ...820 9506 46 ed0821 MRB 3000 SDD SOLAR DIRECT DRIVE ICE LINED BLOOD STORAGE REFRIGERATOR Instructions for Use...

Page 2: ......

Page 3: ...certified according to the safety standard IEC 60335 1 This equipment is designed to be safe at least under the following conditions based on the IEC 60335 1 Indoor use Altitude up to 2000 m Temperat...

Page 4: ...________________15 4 2 Opening and closing___________________________________15 4 3 Important Advice for Storage ___________________________16 4 4 Operating controls __________________________________...

Page 5: ...5 1 Model Overview Group Model name Ice lined Blood Storage Refrigerators MRB3000SDD...

Page 6: ...refrigerator for Solar Direct Drive without battery storage t has a bank of icelining elements installed inside the cabinet During operation these elements freeze and act as cold storage for protectin...

Page 7: ...n that on the nameplate may cause fire or electric shock Do not store flammable or explosive substances inside these appliances e g aerosols with flammable propellant These may cause explosion or fire...

Page 8: ...source The location must be protected against rain and dust This device cannot be used is not for use in facilities at altitudes of 2000m or higher The use of hydrogen peroxide H2O2 VHP for the disinf...

Page 9: ...bmitting the delivery note or proof of purchase Do not operate a unit that has been damaged in transit If you are unsure contact your sales outlet and ask them The unit must be transported in an uprig...

Page 10: ...the ice lined blood storage refrigerators Patients do not come into direct contact with the refrigerators 2 4 3 Intended use environment The Ice lined blood storage refrigerators are stationary type...

Page 11: ...between 2 C and 6 C for no longer than five days prior to use After concentration and irradiation the storage time must be no longer than 24 hours 2 4 5 Contraindications The devices are not intended...

Page 12: ...ctric shock may result due to corrosion The wall socket must be easily accessible When the unit is in its final position it is important to immobilise the unit using the feet CAUTION Optimal function...

Page 13: ...e described in this manual follow the internal policy and procedures Before using the refrigerator it should be cleaned inside and outside For cleaning use only gentle cleaning agents Never use aggres...

Page 14: ...ms with other electrical equipment causing malfunction the unit should be connected to a separate circuit Never connect it to a single socket with other electrical equipment by means of a multi socket...

Page 15: ...ights up when there is a minimum of solar energy available Supposed that there is enough solar energy available after switching on the device starts working after a delay of 5 minutes This also applie...

Page 16: ...nner temperature rises to high Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals correctly against the cabinet Do not bring more than 18 new blood...

Page 17: ...re and the displayed max and min values 4 4 3 Display In normal use the inside temperature of the appliance is displayed in the main area of the diplay C with one decimal max min main area min area ma...

Page 18: ...error E1 4 Warning Display sensor below reading range E1 5 Warning Display sensor above reading range E2 1 Alarm Heating sensor not connected E2 2 Alarm Heating sensor short circuit E2 3 Alarm Heating...

Page 19: ...cessary for the correct daily use of the SDD unit its presence is mandatory When it is not connected an alarm is created When it is damaged detected by a low voltage on the poles another alarm is crea...

Page 20: ...ositioning of solar panels etc Check the actual situation of the panels and if the problem persists after corrective actions notify Refrigerator Technician Thermal cut out of electronic unit E5 5 The...

Page 21: ...the capacity of the battery is no longer sufficient to take over the control function the alarm message E4 2 appears on the display If no battery is connected the system will generate an E4 1 alarm me...

Page 22: ...ly gentle cleaning agents Never use aggressive or caustic cleaning agents scouring powder steel wool abrasive sponges or chemical solvents When cleaning make sure that no fluids of any kind run into t...

Page 23: ...side air to go in and form condensation In such cases the hinges should be adjusted or the gaskets should be changed Restrict and reduce frequency of opening the MRB3000SDD Open only when absolutely n...

Page 24: ...values Check if the ambient temperature is too high outside of the opreating range of the device which can cause the interior temperature too rise Check if there is enough air circulation arround the...

Page 25: ...perational C Max Humididty range Altitude Refrigerant Type Refrigerant quantity g Hold over time at 32 c Hold over time at 43 c Cool down Energy consumption stable running Energy consumption cool down...

Page 26: ...26...

Page 27: ...27...

Page 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...

Reviews: