Application Notes
54
The following six cursor measurements display the set position (independent variable) and measured value (dependent
variables) where the movable cursor intersects the measurement trace.
The position or value measurement text for each dependent variable is displayed in the color of its channel. The
independent variable is white.
Note that the intersection of the movable cursors and the CCDF traces can be moved outside the visible display area.
This does not affect the measurements in any way.
TEXT
TERM
DEFINITION
Cursor Pwr
Cursor Powe
Cursor Mode - Power Ref
Reference
The reference power level in dBr set by the user to define the measure-
ment point on the normalized CCDF for probability in percent.
Cursor Mode - Percent
The measured power level in dBr of the normalized CCDF at the Proba-
bility in percent specified by the user.
Cursor Pct
Cursor Percentage Cursor Mode - Power Ref
The measured probability in percent of the normalized CCDF at the refer-
ence power level specified by the user.
Cursor Mode - Percent
The probability in percent set by the user to define the measurement
point on the normalized CCDF for power level in dBr.
Total
Time
The total time in Hours:Minutes:Seconds that the data acquisition has
been running.
Points
The total number of data samples in MSa that has been acquired for
each channel in the current run.
Table 5.6
Cursor Measurements
Note:
The total number of data samples is affected by the terminal settings. If Terminal Action is set to decimate, then
the sample count will be halved each time the Terminal Count or Time is reached. This should have very little visible
effect on the CCDF values, since the entire population is decimated uniformly. If Terminal Action is set to restart,
then the sample count will be cleared to zero each time the Terminal Count or Time is reached.
5.2 Measurement Accuracy
The measurement accuracy of the RFM3000 is completely contingent upon the USB sensor with which it is being used.
Please reference the sensor datasheet and/or associated uncertainty calculator for measurement uncertainties associated
with a specific sensor.