The RS232 interface does not support flow control. The programmer should be
aware of this limitation and pay attention to the Instrument command process
time. If the remote commands are sent too fast the internal buffer may overrun
and cause communication errors. Therefore, it is mandatory to add a delay
between commands so that the Instrument has sufficient time to process.
13.7
Event_SW
The Event_SW output can be used to control external devices. The Instrument uses a
opto-coupler with an NPN transistor output.
Event switch ratings
Voltage
- 6 to 60 V maximum
Current
50 mA maximum
Table 21 - Event Switch Power
Figure 56 - Event Switch Diagram
To activate this function, connect to pin 13 (E) and pin 25 (Event_SW-), and
then give switch on/off command through SCPI command.
Command
Description
TEST:DIGI ON
TEST:DIGI:IO:SWITCH ON
TEST:DIGI:IO:SWITCH OFF
TEST:DIGI OFF
To activate the switch on/off function
Switch on
Switch off
To disable the switch on/off function
Table 22 - Event Switch Commands
Note: DIGI must be set ON to make changes to the switch on/off state.
Summary of Contents for 9830B Series
Page 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Page 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Page 28: ...15 6 Menu Tree ...
Page 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Page 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Page 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Page 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...