340 Appendix C
Fan Noise
Operating Limits
Disinfection
Dimensions
Electromagnetic Compatibility
Complies with requirements for Class A devices of EN60601-1-2 [2].
Noise
Noise during operation
<30dBA
Environmental Limits
Operating temperature
+10 to +40
°
C (+50 to +104
°
F)
Storage temperature
–25 to +60
°
C (–13 to +140
°
F)
Operating and storage humidity
Max 85% RH non-condensing
Operating pressure
700–1060 hPa (normal atmospheric pressure)
Chemical resistance of console during disinfection
Wiping with
Sodium hypochlorite up to 2% in water
Chlorhexidine gluconate 5% in water
Chlorhexidine 1.5% and cetrimide 15% in water
Chlorhexidine 0.5% in ethanol 70%
Glutaraldehyde 2% solution
Ethanol 70% in water
Isopropanol 70% in water
Approximate Dimensions
Height
1410–1500 mm (adjustable)
Width
530mm
Depth
780mm
Weight
77kg (excluding transducers and printer)
Summary of Contents for Pro Focus 2202
Page 1: ...English BB1279 A June 2005 Pro Focus 2202 Extended User Guide ...
Page 14: ...14 ...
Page 15: ...Part 1 Basics ...
Page 16: ......
Page 32: ...32 Chapter 1 ...
Page 48: ...48 Chapter 2 ...
Page 49: ...Part 2 Working with the Image ...
Page 50: ......
Page 98: ...98 Chapter 5 ...
Page 117: ...Part 3 Imaging Modes ...
Page 118: ......
Page 136: ...136 Chapter 8 ...
Page 152: ...152 Chapter 10 ...
Page 164: ...164 Chapter 12 ...
Page 165: ...Part 4 Setting up and Maintaining Your System ...
Page 166: ......
Page 200: ...200 Chapter 13 ...
Page 208: ...208 Chapter 14 ...
Page 209: ...Part 5 Pro Packages ...
Page 210: ......
Page 288: ...288 Chapter 19 ...
Page 313: ...Part 6 Appendixes ...
Page 314: ......
Page 344: ...344 Appendix C ...