334 Appendix B
To replace the front panel:
1
Remove the sliding cover.
2
Slide the front panel back into the groove.
3
Tighten the black screws.
4
Reconnect to the scanner.
To remove the monitor:
1
Move the monitor base to the top position.
2
Disconnect the monitor from the scanner.
3
Disconnect the internal power supply connector and the USB connector for the
circuit in the monitor base.
4
Remove the sliding cover.
5
Loosen the two black screws on the back of the scanner under the monitor.
6
Lift the monitor base carefully out of the groove.
To replace the monitor:
1
Slide the monitor base into the groove.
2
Tighten the black screws.
3
Reconnect to the scanner.
References
[1] EN 60601–1–1:2001 Medical electrical equipment – Part 1-1: General requirements for
safety. Collateral standard: Safety requirements for medical electrical systems.
Summary of Contents for Pro Focus 2202
Page 1: ...English BB1279 A June 2005 Pro Focus 2202 Extended User Guide ...
Page 14: ...14 ...
Page 15: ...Part 1 Basics ...
Page 16: ......
Page 32: ...32 Chapter 1 ...
Page 48: ...48 Chapter 2 ...
Page 49: ...Part 2 Working with the Image ...
Page 50: ......
Page 98: ...98 Chapter 5 ...
Page 117: ...Part 3 Imaging Modes ...
Page 118: ......
Page 136: ...136 Chapter 8 ...
Page 152: ...152 Chapter 10 ...
Page 164: ...164 Chapter 12 ...
Page 165: ...Part 4 Setting up and Maintaining Your System ...
Page 166: ......
Page 200: ...200 Chapter 13 ...
Page 208: ...208 Chapter 14 ...
Page 209: ...Part 5 Pro Packages ...
Page 210: ......
Page 288: ...288 Chapter 19 ...
Page 313: ...Part 6 Appendixes ...
Page 314: ......
Page 344: ...344 Appendix C ...