138 Chapter 9
Adjusting the Thermal Index Limit
Before you use Color mode, check that the TI settings are appropriate. The current
TI tissue type and limit are displayed just under the B-K Medical logo in the upper
left corner of the screen.
The absolute TI limit for each tissue type is set by the factory to conform to FDA
standards (see “Acoustic Output” on page 26), but you may want to set a lower TI
limit for some purposes. There are 2 types of settings you can vary:
•
Tissue type (TIS – soft tissue, TIC – cranial, TIB – bone)
•
TI limit (not exceeding the factory-set limits)
To select tissue type:
•
Click the TI type displayed and select TIB, TIC or TIS (or point at TI and press
[
+/-
]
).
To adjust the TI limit:
•
Click the limit displayed after the TI type and select the value you want.
The current TI is displayed as TIx:z.z<y.y, where x is S, C, or B; y.y is the limit
you select; and z.z is the actual TI.
Adjusting MI
You can adjust the maximum allowed MI (mechanical index).
To adjust the MI limit:
•
Click MI above General on the left side of the screen and drag the slider.
Turning Color Mode On or Off
When you turn on Color mode, B-mode must be scanning.
To turn on Color mode:
•
Press
[
Color
]
.
The color box appears.
Summary of Contents for Pro Focus 2202
Page 1: ...English BB1279 A June 2005 Pro Focus 2202 Extended User Guide ...
Page 14: ...14 ...
Page 15: ...Part 1 Basics ...
Page 16: ......
Page 32: ...32 Chapter 1 ...
Page 48: ...48 Chapter 2 ...
Page 49: ...Part 2 Working with the Image ...
Page 50: ......
Page 98: ...98 Chapter 5 ...
Page 117: ...Part 3 Imaging Modes ...
Page 118: ......
Page 136: ...136 Chapter 8 ...
Page 152: ...152 Chapter 10 ...
Page 164: ...164 Chapter 12 ...
Page 165: ...Part 4 Setting up and Maintaining Your System ...
Page 166: ......
Page 200: ...200 Chapter 13 ...
Page 208: ...208 Chapter 14 ...
Page 209: ...Part 5 Pro Packages ...
Page 210: ......
Page 288: ...288 Chapter 19 ...
Page 313: ...Part 6 Appendixes ...
Page 314: ......
Page 344: ...344 Appendix C ...