The physician is responsible for ensuring that Stimuplex

®

 HNS 12 is used properly.

1.   Apply skin electrode.
2.   Plug electrode cable into the stimulator and connect skin electrodes.

3.   

Switch on Stimuplex

®

 HNS 12.          Display: 

0.00

;  the basic settings are

displayed in the upper right.

W

. 0

1.

01

.0

8/

   

   

 6

04

35

71

   

   

  0

11

 D

at

of

 la

st

 re

vis

io

n:

 Ja

nu

ar

y 2

01

5

Additional stimulation duration values (ms key):

 0.05 ms / 0.10 ms / 0.30 ms / 0.50ms / 1.00ms

Menu: Setup –> Options –> Set Extra stimulus duration     .

4.   Changing the settings during use:

4.1 Change maximum current (mA key):

 5 mA / 1 mA possible

4.2 Change stimulus duration for 1Hz/2Hz (ms key):

  0.10 ms / 0.30 ms / 1.00 ms for stimulation

(Only framed values can be changed and are immediately active.)

Alternatively: Select the values with the up/down menu keys

–>

Press            5 mA immediately press          again  1 mA

5.   

Plug electrode cable into stimulation needle.

6.   

Set the stimulation current to the desired value by turning dial.

      Red LED blinks; clear warning pitch, warning message: Patient current 0.00

7.   

Perform puncture with stimulation needle. Warning message disappears. Green LED,

      Tone sounds at stimulation frequency.

8.   

Advance the needle until you observe the desired muscle contractions.

9.   

When the desired current is reached, turn the dial to reduce the stimulation current until

      weak contractions are noticeable.
      

Use CAUTION in the lower stimulus current range.

10. 

Inject test dose. Motor responses disappear. Set stimulation current back to maximum.

      (no motor responses should be observed) Inject remaining dose.

11. 

Switch-off:        Hold down key for 1 second (the last basic menu settings are reactivated).

      Stimulator switches off.

4.3 Changing the frequency (Hz key):

 2 Hz / 1Hz / SENSe (if SENSe option enabled)

Press           2Hz    press  1Hz    again –>                      (for SENSe)

Note:
If the circuit is interrupted or the impedance too high: Red light, warning pitch, warning message appear.
Please check: Is skin electrode dry? - Loose? Stimulation needle plugged in? Cable defective?

Hz

Press            0.10 ms press          again   0.30 ms –>

1.00 ms

Press           0.05ms press         immediately again  0.10ms –>

0.30ms  –>  0.50ms  –>  ....

B. Braun Melsungen AG

 · Carl-Braun-Straße 1 · 34212 Melsungen · Germany · Tel +49 (0) 56 61 71-0 · www.bbraun.com

Rx only

ms

ms

ms

ms

ms

ms

ms

mA

mA

Hz

B. Braun Medical Inc. 

· 824 Twelfth Avenue · Bethlehem, PA 18018-3524 · USA · Phone: 1-800-854-6851 · www.bbraunusa.com

frequency 1 Hz and 2 Hz. When using SENSe the pre-defined impulse sequence is

0.10 ms - 0.10 ms - 0.15 to 1.0 ms

6043571_Stimuplex_HNS12_KurzAL_GB.indd   2

04.02.15   09:31