Perfusor® compact, 2.1 gb
8 - 1
8
8 -
Procedural Instructions on the TSC
Visual Inspection
Unit, in General
Completeness, external damage, safe fit of the battery compart-
ment cover and syringe table.
Check cleanliness of device. Check labels and readability.
Syringe Fastening
Check function with OPS 50 ml syringe.
(Syringe holder, axial positioner, drive head, clamp, and push-but-
ton sensor)
Membrane Keyboard
Check adhesion, cleanliness and fit.
Battery Compartment and -Contacts
Check state of contacts (tight fit, not bent).
Unit Feet
Check unit feet for completeness and proper fit.
Mains Lead and Connector
Completeness, damage.
MFC Lead and Connector
Completeness, damage.
Holder for Pole Fixation, Side Snap-in Mechanism
Check function.
8 - 6
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...