Unit Elements
4
4 - 14
Infusomat® fmS 2.3 gb
4.10
Pressure Sensor
Test Equipment
Ord. No.
Pressure calibration device . . . . . . . . . . . . . . . . . . . . . . 0770 5018
Calibration gauge 4 mm (for adjustment after . . . . . . 0770 1489
exchange of the pump cover)
Designation
Ord. No.
Pressure sensor (cmpl.). . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353
(only delivered when a pressure
calibration device is available)
Exchange
CAUTION
The pressure sensor unit is mechanically very sensitive. It must not
be dismounted, which means that the pressure sensor board and
the spring element must not be detached from the pressure sensor
support (plastic part). Make sure that the movement of the coil
core inside the coil is not hindered.
The pressure sensor consists of the light barrier board with EEP-
ROM and the pressure sensor board with holder. Coil core and
bending element are screwed to this holder. Both boards are con-
nected via a flat cable.
1.
Dismount rear panel
or pump
unit
2.
Pull off the connecting cable to the microprocessor board.
3.
Loosen and remove screw (1) of the light barrier board.
4.
Loosen screw (2) of the pressure sensor and remove it togeth-
er with plain washer and serrated lock washer. Then remove
pressure sensor with pressure sensor board and light barrier
board.
5.
Fit pressure sensor with pressure sensor- and light barrier
board.
Note
Make sure that the pressure sensor slide can smoothly run in the
guides and that the coil core can move inside the coil.
6.
Calibrate pressure sensor:
a)
Switch on unit and call in the pressure sensor reading in
(see „Pressure Sensor Function
b)
Loosen screw (2) on the side of the pressure sensor unit
(2.5 mm Allen key).
Fig.: 4 - 14
Pressure sensor with
Light barrier
Screw (1)
Screw (2)
board
pressure sensor board
Serrated lock washer
Plain
washer
Do not loosen screw
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...