Infusomat
®
fmS
Instructions for Use
Software IFME/IFMe
Page 1: ...Infusomat fmS Instructions for Use Software IFME IFMe...
Page 2: ...urer s information for possible incompatibilities of equipment resp drugs Use only compatible combinations of equipment accessories working parts and disposables It is recommended to use original Infu...
Page 3: ...Infusomat fmS is according to IEC EN 60601 1 resp IEC EN 60601 2 24 a volumetric infusion pump for infusion of small and large volumes at highest accuracy and is suitable for intravenous and intra ar...
Page 4: ...ss the black button Overview Display All important information in plain text Green background illumination only if connected to mains or a key is pushed at battery operation Correct Input CLEAR Displa...
Page 5: ...f mains power failure the pump switches to battery operation Battery operation time 3 5 h at highest delivery rate Automatic over load protection Mains fuse directly above the mains power connector Po...
Page 6: ...essary use an appropriate adapter 4 Switch On with Green mains power control or yellow battery control alarm tone and display LED s are briefly activated 5 Puncture 6 Setting the Delivery Rate In the...
Page 7: ...ed residual volume with C in display appears target volume ml and confirm with VOL If desired enter new target volume To continue infusion the target volume has to be displayed as ml or a new target v...
Page 8: ...evice alarms At the moment of an event also the volume infused from switching on is recorded The history function is activated ex works After a software update the function has to be installed again s...
Page 9: ...keys are pressed Per ml bolus administered a short audible signal sounds Interval bolus Automatic bolus in set time interval In case of manual bolus administration the interval bolus is skipped Selec...
Page 10: ...s necessary for drop control to appear Press key below OFF no drop control appears To switch on again Press key below ON Exit from selection Press key below END Check Battery Capacity Remaining batter...
Page 11: ...ired quantity and time unit Enter values at flashing star as described and confirm with OK Automatically calculated values rate or dosage are displayed First check rate displayed for plausibility then...
Page 12: ...primary rate and volume and confirm The softkey PIGY only will be displayed after a target volume had been entered before Press key below PIGY enter piggyback rate and volume and confirm Start infusio...
Page 13: ...ssing key The alarm continues in short intervals until the bat tery is fully discharged Battery Alarm Alternately rate and AAA A are displayed The text display shows battery discharged connect to main...
Page 14: ...tteries Ageing After 2 years the original capacity is only approx 50 The lifetime of the battery can be extended by completely discharging from time to time and recharging afterwards Interface Interfa...
Page 15: ...deviation 2 5 11 19 p t min 31 Epmax Epmin 0 The graphs show the accuracy uniformity of flow in relation to time Allow for the following The delivery behaviour resp delivery accuracy is essentially i...
Page 16: ...ted at appliance inlet External extra low voltage 12 V DC Staff call Max 24 V 1 A 24 VA Arbitrary connection polarity VDE 0834 EMC EN 55011 IEC EN 60601 1 2 and IEC EN 60601 2 24 Time of operation 100...
Page 17: ...tion 0 1 9999 9 ml in 0 1 ml increments Occlusion alarm pressures Alarm reaction time Max bolus volume measured at 22 C with OIL infusion set Mechanical occlusion pressure limit Occlusion alarm pressu...
Page 18: ...chnical Safety Check Service The Technical Safety Check is recommended to be carried out every 2 years and should be documented Servicing work must be carried out exclusively by personnel instructed b...
Page 19: ...870 1644 Original Infusomat infusion set CVP with line for CVP measuring 340 cm 870 0010 Original Infusomat infusion line 250 cm 870 0036 Original Infusomat infusion set 5 m with 5 m filter 275 cm 87...
Page 20: ...erial No 3891 2341 Drawing No M671010100F04 05 01 Printed on pulp bleached 100 chlorine free B Braun Melsungen AG P O Box 11 20 D 34209 Melsungen Tel 49 0 56 61 71 0 Fax 49 0 56 61 71 20 44 HOSPITAL C...