Unit Elements
4
4 - 2
Infusomat® fmS 2.3 gb
4.3
Door Lock
Designation
Ord. No.
Door lock complete with push button . . . . . . . . . . . . . 3450 5601
Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440
Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790
Exchange
1.
Open door and unlatch the spring holder.
2.
Remove countersunk screw and press out the mounting by
pressing the holder for the door lock.
3.
Remove door lock in an upward direction and exchange.
4.
Assembly is done in reverse order.
Check
Pump unit check (only mechanical occlusion pressure).
Fig.: 4 - 2
Spring holder
S M 3x6