Azbil Corporation
Maintenance and troubleshooting of this instrument
Model: MVC10A/10F - Multivariable Air Flow Meter 5-5
5-2:
Troubleshooting
Types of troubles
The following three types of troubles are conceivable when activating and starting
this instrument:
Such troubles that occur when the actual service conditions differ from the
specifications of this instrument
Troubles caused by the setting and operating errors
Troubles due to the failure of this instrument
If any trouble occurs, follow the troubleshooting guide described here and take proper
actions.
Troubles at the start of operation
Troubleshooting
If any trouble occurs, take remedial actions in accordance with the following
tabulated instructions. If the trouble cannot be solved even after taking these actions,
this instrument may be malfunctioning. Then, please contact our sales representatives
at our branch, sales office, or our distributors where you purchased this instrument, or
our customer services department of our head office.
Trouble
Check points and remedies
Nothing appears on the display after
turning on the power switch.
• Ensure that the power supply voltage is correctly applied.
• Ensure that power supply cable is connected.
• Ensure that the fuse is not blown.
No output is transmitted even if the
power is turned on.
• Ensure that the signal lines are correctly connected.
No pulse output is given.
• Ensure that the pulse output lines are correctly wired.
Indication of flow rate remains
unchanged at the zero point.
• Ensure that the details of settings are correct.
•
Ensure that the flow rate is not deviating from the low
flow cut range.
• Ensure that the piping is not clogged.
While the flow rate should be zero, a
value other than zero is indicated.
•
Even a minimal deviation from the zero point may be
indicated in a large value in the scale of
¡
/day and
¡
/
hour. If this is the case, either change the indication unit
or conduct the zero adjustment or set a low flow cut
value.
• Ensure that the polarity of detector terminals and
connection of the converter are correct.